Specifications and Model Number

Note 1: Customer must supply tagging information; Up to 3 lines allowed. (22 characters for each line)

Note 2: This option for use with HPW7000 assembly only.

Note 3: Standard range should be selected if measuring the following conductivity, resistivity or TDS ranges:

STANDARD RANGE

Conductivity

TDS

Cell Constant

0-1.999

0-1999 ppb

0.01

0-19.99

0-19,990 ppb

0.1

0-199.9

0-199.9 ppm

1.0

0-1999

0-1999 ppm

10

0-19,990

0-10,000 ppm

25 or 50

Resistivity

 

Cell Constant

0-19.99M

 

0.01

0-1.999M

 

0.1

0-199.9k

 

1.0

0-19.99k

 

10

0-1.999k

 

25 or 50

Note 4: Wide range should be selected if measuring the following

conductivity, TDS or weight % concentration ranges:

WIDE RANGE

Conductivity

 

TDS

Cell Constant

100-1999

 

0-1999 ppm

0.1

1000-19,990

 

0-19,990 ppm

1.0

10-199.9 mS

 

0-199.9 ppt

10

10-500 mS

 

 

25

20-1000 mS

 

 

50

Weight %

Concentration

Cell Constant

0-18.00% Hydrochloric Acid

50

0-19.99% Sulfuric Acid

 

50

0-19.99% Sodium Hydroxide

25 or 50

0-19.99% Sodium Chloride

25 or 50

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

2-9

Page 41
Image 41
Honeywell 9782 Series manual Standard Range