Maintenance

Step

Screen

 

 

8OUTPUT1 SPAN OUTPUT1 ZERO OUTPUT2 SPAN OUTPUT2 ZERO OUTPUT3 SPAN OUTPUT3 ZERO

DIAGNOSTICS

9SPAN CAL OUTPUT1

USE UP/DOWN ARROW

KEYS TO ADJUST

ENTER TO SAVE

DIAGNOSTICS

10OUTPUT CALIBRATION

SAVED

MENU TO EXIT

DIAGNOSTICS

11

12

13

14

Action

The first display will reflect the quantity of analog outputs supported by the hardware.

Select the value to be calibrated.

Go to the next screen.

Use the UP and DOWN keys to correct the reading on the test meter.

To save the correction press ENTER.

This screen confirms that the value was saved.

To calibrate additional values, repeat Steps 1 through 10, including powering down the unit before changing the connections to the output terminals.

When output calibration has been completed, re-install field wiring removed in Step 4. Disconnect power before opening

case.

Replace the safety cover and secure it with the screw.

Close the case and power up the unit. Do not apply power

until the case is closed.

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

10-5

Page 121
Image 121
Honeywell 9782 Series manual Saved, Until the case is closed