Honeywell 9782 Series manual Action Go to Maintenance Lead Resistance Comp

Models: 9782 Series

1 148
Download 148 pages 17.14 Kb
Page 129
Image 129

Maintenance

Procedure

To use the 9782’s automatic compensation for lead resistance, enter the necessary values using the procedure in Table 10-7.

Table 10-7 Procedure for Entering Values for Lead Resistance Compensation

Step

1

2

Screen

MENU

CONFIGURATION

CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS

PREV NEXT

MAINTENANCE

OFFLINE FUNCTIONS

INSTRUMENT SETUP

LEAD RESISTANCE COMP

DIAGNOSTICS

PREV NEXT

LEAD RESISTANCE COMP

 

WIRE GAUGE

12AWG

 

 

WIRE LENGTH

 

 

 

CELL1

100FT

 

 

CELL2

150FT

 

 

DIAGNOSTICS

 

 

 

PREV

NEXT

 

 

 

 

 

Action

Go to MAINTENANCE LEAD RESISTANCE COMP.

Use this screen to specify the gauge used for leads to both cells, and the length of one lead to each cell.

If mixed gauges are used, or the length exceeds 1999 feet, refer to the Introduction of this sub-section for guidance in computing the values to enter here.

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

10-13