Conductivity/Resistivity Analyzer/Controller

WARNING

While the unit is powered, a potentially lethal shock hazard exists inside the case. Do not open the case while the unit is powered. More than one switch may be required to disconnect power.

Step

1

2

3

4

5

Table 6-4 Procedure for Wiring Relays

Action

If you have not already noted the relay assignments made during I/IO setup, go to I/O SETUP on the main menu to view the display showing relay output use. Note the assignments shown. You must wire the unit to match these assignments in order for the unit to work as expected. (See Section 5.)

Turn off the power to the Analyzer/Controller. More than one switch may be required to remove power.

With the power off, open the case:

Grasp the bottom center portion of the front bezel and pull it downward and toward you slightly to disengage the bottom of the bezel from the edge of the case.

Lift the bezel gently to disengage it from the top edge of the case.

Swing the bezel to the left. (The bezel and display assembly is mounted on pivot arms.) The safety cover with its warning label will be visible.

With the power off remove the safety cover by removing the single screw holding it in place.

Wire the relays in accordance with the system setup assignments. Refer to the wiring diagram provided with the external device receiving the output and to Figure 6-1.

WARNING

An external disconnect switch is required for any hazardous voltage connections to the relay outputs.

6

7

Replace the safety cover and secure it with the screw.

Close the case and power up the unit. Do not apply power until the case is closed.

6-8

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

Page 90
Image 90
Honeywell 9782 Series manual Conductivity/Resistivity Analyzer/Controller