Conductivity/Resistivity Analyzer/Controller

Table 5-4 Procedure for Configuring PFT Control

Step

1

Screen

MENU

CONFIGURATION

CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS

PREV NEXT

Action

Go to CONFIGURATION DISCRETE CONTROL PFT.

2

3

CONFIGURATION

CAT/RETRANSMISSION

ALARMS

DISCRETE CONTROL

ADVANCED FEATURES

DIAGNOSTICS

PREV NEXT

DISCRETE CONTROL

ON/OFF

PFT

DAT

DIAGNOSTICS

PFT CONTROL

DB FREQ RATE 100 P/M

PFT1

PFT2

DIAGNOSTICS

PFT 1

 

OUTPUT

RLY2

 

 

INPUT

 

C1

 

 

SETPOINT

15000

µS

 

 

PB LIMIT

250

µS

 

 

 

 

 

 

DIAGNOSTICS

 

 

 

 

 

 

 

 

 

 

ATTENTION: Any value entered for one of these control- related parameters will take effect as soon as it is entered.

Specify the maximum frequency rate (in pulses per minute) representing 100% output. (Be sure this rate is compatible with the controlled device.) This rate will be used for all PFT outputs in the Analyzer/Controller.

Use this screen to select the PFT control output to be configured.

The relay assigned to this PFT control output during I/O setup will already be filled in. (This is read-only information.)

Specify the process value or computed value to be used as the input.

Enter the setpoint at which the output should be turned on.

Enter the PB Limit. This is the value for which the output’s pulse output rate will be at its maximum frequency.

5-14

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

Page 70
Image 70
Honeywell 9782 Series Procedure for Configuring PFT Control, Action Go to Configuration Discrete Control PFT, Pft, PFT1