Conductivity/Resistivity Analyzer/Controller
Table 5-4 Procedure for Configuring PFT Control
Step
1
Screen
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
PREV NEXT
Action
Go to CONFIGURATION DISCRETE CONTROL PFT.
2
3
CONFIGURATION
CAT/RETRANSMISSION
ALARMS
DISCRETE CONTROL
ADVANCED FEATURES
DIAGNOSTICS
PREV NEXT
DISCRETE CONTROL
ON/OFF
PFT
DAT
DIAGNOSTICS
PFT CONTROL
DB FREQ RATE 100 P/M
PFT1
PFT2
DIAGNOSTICS
PFT 1
| OUTPUT | RLY2 |
| |
| INPUT |
| C1 |
|
| SETPOINT | 15000 | µS |
|
| PB LIMIT | 250 | µS |
|
|
|
|
| |
| DIAGNOSTICS |
|
| |
|
|
| ||
|
|
|
|
|
ATTENTION: Any value entered for one of these control- related parameters will take effect as soon as it is entered.
Specify the maximum frequency rate (in pulses per minute) representing 100% output. (Be sure this rate is compatible with the controlled device.) This rate will be used for all PFT outputs in the Analyzer/Controller.
Use this screen to select the PFT control output to be configured.
The relay assigned to this PFT control output during I/O setup will already be filled in. (This is
Specify the process value or computed value to be used as the input.
Enter the setpoint at which the output should be turned on.
Enter the PB Limit. This is the value for which the output’s pulse output rate will be at its maximum frequency.
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual | 7/99 |