Calibration
Next, perform the calibration. Once calibration has been completed, go back to I/O setup and set the solution temperature compensation type to the TDS choice, for example “NaCl (TDS)”.
At this point you are ready to enter the TDS conversion factor as described in Table
Procedure
To enter the TDS conversion factor follow the instructions in in Table
Table 8-5 Procedure for Entering TDS Conversion Factor
Step
1
Screen
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
CALIBRATION
CELL CAL FACTOR
TDS CVSN FACTOR C1
TDS CVSN FACTOR C2 REMOVE CAL TRIM C1 REMOVE CAL TRIM C2
DIAGNOSTICS
Action
Go to CALIBRATION TDS CVSN FACTOR C1.
If you plan to calibrate, do it before entering the theoretically derived TDS conversion factor (see previous page).
2
|
|
|
|
|
| TDS CONVRSN FACTOR |
| Enter the appropriate TDS conversion factor for the | |
|
|
|
| |
| C1 | 0.5 |
| selected cell. |
|
|
|
|
|
| DIAGNOSTICS |
|
|
|
| |
|
| |
|
|
|
|
|
3
4
TDS CONVRSN FACTOR
ACCEPTED AND SAVED
DIAGNOSTICS
CALIBRATION
CELL CAL FACTOR TDS CVSN FACTOR C1
TDS CVSN FACTOR C2
REMOVE CAL TRIM C1 REMOVE CAL TRIM C2
DIAGNOSTICS
If the factor entered is within the permitted range for the cell constant (see Table
If the value is
Once an acceptable factor has been entered for Cell 1, repeat the process for Cell 2.
7/99 | 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual |