Calibration

Next, perform the calibration. Once calibration has been completed, go back to I/O setup and set the solution temperature compensation type to the TDS choice, for example “NaCl (TDS)”.

At this point you are ready to enter the TDS conversion factor as described in Table 8-5.

Procedure

To enter the TDS conversion factor follow the instructions in in Table 8-5.

Table 8-5 Procedure for Entering TDS Conversion Factor

Step

1

Screen

MENU

CONFIGURATION

CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS

CALIBRATION

CELL CAL FACTOR

TDS CVSN FACTOR C1

TDS CVSN FACTOR C2 REMOVE CAL TRIM C1 REMOVE CAL TRIM C2

DIAGNOSTICS

Action

Go to CALIBRATION TDS CVSN FACTOR C1.

If you plan to calibrate, do it before entering the theoretically derived TDS conversion factor (see previous page).

2

 

 

 

 

 

 

TDS CONVRSN FACTOR

 

Enter the appropriate TDS conversion factor for the

 

 

 

 

 

C1

0.5

 

selected cell.

 

 

 

 

 

 

DIAGNOSTICS

 

 

 

 

<--

-->

 

 

 

 

 

 

 

3

4

TDS CONVRSN FACTOR

ACCEPTED AND SAVED

DIAGNOSTICS

CALIBRATION

CELL CAL FACTOR TDS CVSN FACTOR C1

TDS CVSN FACTOR C2

REMOVE CAL TRIM C1 REMOVE CAL TRIM C2

DIAGNOSTICS

If the factor entered is within the permitted range for the cell constant (see Table 5-10), the Analyzer/Controller will use the entered value.

If the value is out-of-range, the Analyzer/Controller will force the value to the appropriate range limit.

Once an acceptable factor has been entered for Cell 1, repeat the process for Cell 2.

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

8-9

Page 107