Maintenance

10.4Viewing Product Information and Changing Model Number Stored in Memory

Introduction

During I/O setup (see Section 5) the software will display prompts only for those features the hardware can support. To do this, the software uses the model number of the unit stored in memory at the factory. Each combination of features has a unique model number (see Section 2). Therefore, if you add an option card to the 9782, it is very important that you change the model number stored in memory using the procedure described below.

ATTENTION

A special password is required to change the stored model number. If the password was not supplied with the new option card, call the Honeywell Technical Assistance Center.

Before calling for technical support, follow the procedure in Table 10-4 to determine the version of the software in the Analyzer/Controller. The support team will ask for this information.

Procedure

Use the procedure in Table 10-4 to see the software version and model number, and to change the model number if required.

Step

1

Table 10-3 Procedure for Viewing Product Information and

Changing the Stored Model Number

 

Screen

Action

 

 

 

 

 

 

 

 

 

Go to MAINTENANCE INSTRUMENT SETUP

 

MENU

 

 

 

CONFIGURATION

 

 

 

 

 

PRODUCT INFO.

 

CALIBRATION

 

 

 

MAINTENANCE

 

 

 

 

I/O SETUP

 

 

 

 

 

 

 

 

 

DIAGNOSTICS

 

 

 

 

PREV

NEXT

 

 

 

 

 

 

 

 

 

 

 

 

 

MAINTENANCE

 

 

 

 

OFFLINE FUNCTIONS

 

 

 

INSTRUMENT SETUP

 

 

 

LEAD RESISTANCE COMP

 

 

 

 

 

 

 

 

DIAGNOSTICS

 

 

 

 

PREV

NEXT

 

 

 

 

 

 

 

INSTRUMENT SETUP

INSTURMENT TAGGING

PRODUCT INFO

SCREEN CONTRAST

SECURITY

RESET UNIT

DIAGNOSTICS

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

10-7

Page 123
Image 123
Honeywell 9782 Series manual Product Info