Conductivity/Resistivity Analyzer/Controller
Table
Step | Screen |
|
|
1MENU
CONFIGURATION
CALIBRATION MAINTENANCE I/O SETUP
DIAGNOSTICS
2CALIBRATION
CELL CAL FACTOR
CAL TRIM 1
CAL TRIM 2
REMOVE CAL TRIM C1
REMOVE CAL TRIM C2
DIAGNOSTICS
3CAL TRIM
PUT C1 IN
REFERENCE SOLUTION
ATTENTION
WAIT FOR STABLE
READING!
DIAGNOSTICS
PREV NEXT
4CAL TRIM C1
147.00µS
DIAGNOSTICS
5CAL TRIM C1
147.00µS
CAL ADJ FAILED
Action
Prepare container of the reference solution.
Go to CALIBRATION CAL TRIM C 1.
Remove the cell from the process.
Rinse the cell thoroughly with
Immerse it in the reference solution.
Go to the next screen.
The display will show the conductivity of the reference solution as measured by the cell and Analyzer/Controller system.
Once the reading is stable, change the value to match the actual conductivity of the reference solution at its current
temperature.
Press ENTER to save the calibration trim value. If the calibration trim adjustment is successful, the calibration menu will again be displayed.
Return the cell to the process and go to Step 6.
If the attempted calibration trim is greater than the allowable adjustment limit, the operation will not be successful. The CAL TRIM display will remain on the screen and an error message will be displayed on the alarm stripe.
Check the cell and replace if necessary.
6
After successful calibration, repeat the operation for the other cell. When calibration trim has been completed for both cells, return to the main menu.
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual | 7/99 |