Conductivity/Resistivity Analyzer/Controller

Table 8-3 Procedure for Performing Calibration Trim Using a Reference Solution

Step

Screen

 

 

1MENU

CONFIGURATION

CALIBRATION MAINTENANCE I/O SETUP

DIAGNOSTICS

2CALIBRATION

CELL CAL FACTOR

CAL TRIM 1

CAL TRIM 2

REMOVE CAL TRIM C1

REMOVE CAL TRIM C2

DIAGNOSTICS

3CAL TRIM

PUT C1 IN

REFERENCE SOLUTION

ATTENTION

WAIT FOR STABLE

READING!

DIAGNOSTICS

PREV NEXT

4CAL TRIM C1

147.00µS

DIAGNOSTICS

5CAL TRIM C1

147.00µS

CAL ADJ FAILED

Action

Prepare container of the reference solution.

Go to CALIBRATION CAL TRIM C 1.

Remove the cell from the process.

Rinse the cell thoroughly with de-ionized water.

Immerse it in the reference solution.

Go to the next screen.

The display will show the conductivity of the reference solution as measured by the cell and Analyzer/Controller system.

Once the reading is stable, change the value to match the actual conductivity of the reference solution at its current

temperature.

Press ENTER to save the calibration trim value. If the calibration trim adjustment is successful, the calibration menu will again be displayed.

Return the cell to the process and go to Step 6.

If the attempted calibration trim is greater than the allowable adjustment limit, the operation will not be successful. The CAL TRIM display will remain on the screen and an error message will be displayed on the alarm stripe.

Check the cell and replace if necessary.

6

After successful calibration, repeat the operation for the other cell. When calibration trim has been completed for both cells, return to the main menu.

8-6

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99

Page 104
Image 104
Honeywell 9782 Series manual Go to Calibration CAL Trim C, Temperature