I/O Setup and System Configuration

5.7 Setting the Clock

Purpose

A display showing the current time can be accessed when the unit is online. The 24-hour clock must be set before you can use the AutoClean feature described in Sub-Section 5.6.

Procedure

Follow the procedure in Table 5-7 to set the 24-hour clock.

Table 5-7 Procedure for Setting the Clock

Step

1

2

Screen

MENU

CONFIGURATION

CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS

PREV NEXT

CONFIGURATION CAT/RETRANSMISSION ALARMS

DISCRETE CONTROL

ADVANCED FEATURES

DIAGNOSTICS

PREV NEXT

ADVANCED FEATURES

CLOCK SETUP

COMPUTED VALUE

AUTO CLEAN

DIAGNOSTICS

24 HOUR CLOCK SETUP

CLOCKOFF

HOURS13

MINUTES30

DIAGNOSTICS

NEXT

Action

Go to CONFIGURATION ADVANCED FEATURES

CLOCK SETUP.

Set the clock. For example, if you will start the clock at 1:30 PM, enter 13 hours, 30 minutes.

Start the clock and specify what should happen to the clock if the power fails.

ON W/RESET starts the clock now. Following power failure, the clock will be reset to 00:00 and the diagnostic message “RESET CLOCK” will be displayed.

ON W/O RESET starts the clock now. Following power

failure the clock will remain off, and the diagnostic message “RESET CLOCK” will be displayed.

Display of ”RESET CLOCK” will continue until you either reset the clock or turn it off.

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

5-17