I/O Setup and System Configuration
PARAMETER
configuration: retransmission 0%
configuration: retransmission 100%
configuration: CAT control setpoint value
configuration: alarm setpoint value
configuration: configuration: On/Off control cycle timer setpoint value
configuration: PFT control setpoint value
configuration: DAT control setpoint value
configuration: CAT control proportional band limit
configuration: alarm deadband
configuration: On/Off control cycle timer deadband
configuration: PFT control proportional band limit
configuration: DAT control proportional band limit
configuration: alarm delay
configuration: process variable or computed value to be used as basis for discrete control output
configuration: process variable or computed value alarmed
configuration: alarm action
configuration: On/Off control cycle timer period
configuration: On/Off control cycle timer “on” time
configuration: PFT control maximum pulse frequency
configuration: DAT control cycle period
configuration:
configuration:
configuration:
configuration: AutoClean timer set day
configuration: AutoClean timer set hour
configuration: AutoClean timer set minutes
CHOICES or RANGE (Default is underlined)
range depends on measurement type and cell constant; default is value of lower range limit
range depends on measurement type and cell constant; default is value of upper range limit
range depends on measurement type and cell constant; default is value of upper range limit
0 to 100 seconds, default = 0
C1 [Cell 1 value]
C2 [Cell 2 value]
TMP1 [temperature value from Cell 1] TMP2 [temperature value from Cell 2]
T1/2 [temperature 1 and/or 2 - only available for alarming]
Cell1
RATO [computed ratio: Cell2 ]
| (1 − | Cell1 | )x100 | |||
|
|
| ||||
%REJ [computed percent rejection: |
|
| Cell2 | ] | ||
|
| Cell1 | x100 |
| ||
%PSG [computed percent passage: Cell2 | ] |
DIFF [computed difference: Cell1 − Cell2 ]
CO2 [computed parts per million carbon dioxide -
HIGH
LOW
0 to 1000 seconds, default = 60
0 to 100% at 1% resolution, default = 100
1 to 200 pulses per minute, default = 60
0 to 200 seconds, default = 60
OFF
ON W/ RESET
ON W/O RESET
0 to 24, default = 1
0 to 59, default = 0
0 to 28, default = 0
0 to 24, default = 1
0 to 59, default = 0
7/99 | 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual |