Conductivity/Resistivity Analyzer/Controller
PARAMETER
CHOICES or RANGE (Default is underlined)
configuration: AutoClean current day
configuration: AutoClean rinse duration
configuration: AutoClean resume delay
configuration: computed value selection
maintenance: display resolution
maintenance: security enabled
maintenance: password
maintenance: lead resistance compensation wire gauge (AWG)
wide range models only
maintenance: lead resistance compensation length wide range models only
calibration: TDS conversion factor
0 to 28, default = 0
1 to 1999 seconds, default = 60
NONE
Cell1
RATIO [computed ratio: Cell2 ]
DIFFERENCE [computed difference: Cell1 − Cell2 ]
Cell1 x100
%PASSAGE [computed percent passage: Cell2 ] %REJECTION [computed percent rejection
(1 −
:]
PPMCO2 [computed parts per million carbon dioxide -
LOW
HIGH
YES
NO
0000 to 9999; default = 0000
12
14
16
18
1 to 1500 feet default = 1
standard range models with cell constant of 1, 10, 25, or 50:
0.01 to 1.999; default = 0.5 ppm
standard range models with cell constant of 0.01 or 0.1: 10 to 1999; default = 500 ppb
wide range models for all cell constants: 0.01 to 1.999; default = 0.5 ppm
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual | 7/99 |