Diagnostics and Messages

9.5 Off-Line Diagnostics

Introduction

The following off-line diagnostics can be initiated by the operator:

display test (see 9.5.1)

keyboard test (see 9.5.1)

output tests (see 9.5.2)

9.5.1Display and Keyboard Test

Procedure

The procedure for initiating the display and keyboard test is described in Table 9-4.

Table 9-4 Procedure For Initiating Display and Keyboard Tests

Step

1

2

Screen

MENU

CONFIGURATION

CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS

PREV NEXT

MAINTENANCE

OFFLINE FUNCTIONS

INSTRUMENT SETUP LEAD RESISTANCE COMP

DIAGNOSTICS

PREV NEXT

OFF-LINE FUNCTIONS

DISPLAY TEST

KEYBOARD TEST OUTPUT/RELAY TEST OUTPUT CALIBRATON

DIAGNOSTICS

PREV NEXT

DISPLAY TEST

DIAGNOSTICS

DTEST

DISPLAY TEST

F1=TOGGLE MENU=ABORT

DIAGNOSTICS

Action

Go to MAINTENANCE OFF-LINE FUNCTIONS

DISPLAY TEST.

Pressing the specified function key will display a test pattern on the first line of the display. Press the key repeatedly to test the rest of the display.

After confirming that each line of the display is functional, press the MENU key to return to the off-line functions menu.

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

9-5

Page 113