Maintenance
10.7Resetting All Configuration and Calibration Values to Factory Settings
Introduction
A “cold reset” function is available to return ALL configuration values to the factory settings shown in Table
This “cold reset” contrasts with the “warm reset” that occurs following loss of power and at the end of system setup (see Section 5). During a warm reset, alarms are not cleared, and outputs are maintained at their last value prior to the loss of power or initiation of the cold reset. A warm reset has no effect on configuration or calibration values.
Procedure
To do a cold reset, returning all configuration and calibration values to their factory settings, follow the procedure in Table
Table
Step
1
2
Screen
MENU
CONFIGURATION
CALIBRATION
MAINTENANCE
I/O SETUP
DIAGNOSTICS
PREV NEXT
MAINTENANCE OFFLINE FUNCTIONS
INSTRUMENT SETUP
LEAD RESISTANCE COMP
DIAGNOSTICS
PREV NEXT
INSTRUMENT SETUP INSTURMENT TAGGING PRODUCT INFO SCREEN CONTRAST SECURITY
RESET UNIT
DIAGNOSTICS
UNIT RESET
ATTENTION
ALL PARAMETERS
WILL BE RESET TO FACTORY DEFAULTS!
DIAGNOSTICS
RESET
Action
Go to MAINTENANCE INSTRUMENT SETUP RESET UNIT.
Pressing the specified function key at this point will perform a cold reset as described above.
As the reset is completed. the display will go blank, then for a few seconds product information will be displayed. Finally, the display will return to showing process values.
7/99 | 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual |