Maintenance

10.7Resetting All Configuration and Calibration Values to Factory Settings

Introduction

A “cold reset” function is available to return ALL configuration values to the factory settings shown in Table 5-1. During a cold reset the cal factor will be changed to match the cell constant. The calibration trim or TDS conversion factor will be removed. All alarms will be cleared and outputs will go to zero. Some I/O setup values will also be returned to the factory defaults.

This “cold reset” contrasts with the “warm reset” that occurs following loss of power and at the end of system setup (see Section 5). During a warm reset, alarms are not cleared, and outputs are maintained at their last value prior to the loss of power or initiation of the cold reset. A warm reset has no effect on configuration or calibration values.

Procedure

To do a cold reset, returning all configuration and calibration values to their factory settings, follow the procedure in Table 10-6.

Table 10-6 Procedure for Resetting All Configuration to Factory Settings

Step

1

2

Screen

MENU

CONFIGURATION

CALIBRATION

MAINTENANCE

I/O SETUP

DIAGNOSTICS

PREV NEXT

MAINTENANCE OFFLINE FUNCTIONS

INSTRUMENT SETUP

LEAD RESISTANCE COMP

DIAGNOSTICS

PREV NEXT

INSTRUMENT SETUP INSTURMENT TAGGING PRODUCT INFO SCREEN CONTRAST SECURITY

RESET UNIT

DIAGNOSTICS

UNIT RESET

ATTENTION

ALL PARAMETERS

WILL BE RESET TO FACTORY DEFAULTS!

DIAGNOSTICS

RESET

Action

Go to MAINTENANCE INSTRUMENT SETUP RESET UNIT.

Pressing the specified function key at this point will perform a cold reset as described above.

As the reset is completed. the display will go blank, then for a few seconds product information will be displayed. Finally, the display will return to showing process values.

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

10-11

Page 127
Image 127
Honeywell 9782 Series manual Action Go to Maintenance Instrument Setup Reset Unit