Conductivity/Resistivity Analyzer/Controller

Step

Screen

 

 

2PRODUCT INFORMATION

SOFTWARE VER: A1.0

MODEL NUMBER:

9782X-nn-Xn-XXXX

DIAGNOSTICS

EDIT

3FACTORY PASSWORD

OOOO

ENTER PASSWORD

DIAGNOSTICS

4MODEL NUMBER:

9782X-nn-Xn-XXXX

DIAGNOSTICS

Action

The software version is read-only information.

If the model number displayed no longer matches the hardware because you have added or changed an option card, use the specified function key to edit the model number. (See Section 2 for available model number selections.)

Note that a special password is required to change the model number. (This is not the same as the security password described in 10.7.) Enter it when prompted.

This password is included with the instructions for installing the new card.1

ATTENTION: Changing the model number stored in memory will do a “cold reset” of the unit, returning ALL configuration values to the factory settings shown in Table 5-1. During a cold reset some I/O setup values will also be returned to the defaults, and the cal factor will be reset to match the cell constant. The calibration trim or TDS conversion factor will be removed.

The “X” values in the model number are “don’t care” conditions, because these characters do not affect I/O setup and configuration parameters.

Once you have completed the model number change, press ENTER to accept the change.

To cancel the edit and avoid resetting the unit, press the MENU key.

1If you do not have the factory password, call the Technical Assistance Center at 1-800-423-9883.

10-8

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

7/99