Conductivity/Resistivity Analyzer/Controller
Step | Screen |
|
|
2PRODUCT INFORMATION
SOFTWARE VER: A1.0
MODEL NUMBER:
DIAGNOSTICS
EDIT |
3FACTORY PASSWORD
OOOO
ENTER PASSWORD
DIAGNOSTICS
4MODEL NUMBER:
DIAGNOSTICS
Action
The software version is
If the model number displayed no longer matches the hardware because you have added or changed an option card, use the specified function key to edit the model number. (See Section 2 for available model number selections.)
Note that a special password is required to change the model number. (This is not the same as the security password described in 10.7.) Enter it when prompted.
This password is included with the instructions for installing the new card.1
ATTENTION: Changing the model number stored in memory will do a “cold reset” of the unit, returning ALL configuration values to the factory settings shown in Table
The “X” values in the model number are “don’t care” conditions, because these characters do not affect I/O setup and configuration parameters.
Once you have completed the model number change, press ENTER to accept the change.
To cancel the edit and avoid resetting the unit, press the MENU key.
1If you do not have the factory password, call the Technical Assistance Center at
9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual | 7/99 |