I/O Setup and System Configuration

Step

Screen

 

 

13SOLUTION TEMP COMP

C1

ACID/NH3/ETA

C2

NACL

DIAGNOSTICS

PREV NEXT

14ANALOG OUTPUTS

OUTPUT1 C1

OUTPUT2 C2

OUTPUT3 RATO

DIAGNOSTICS

PREV NEXT

15OUTPUT SCALING OUTPUT1 LINFIXEDRNG

OUTPUT2 LINAUTORNG

OUTPUT3 BI-LINEAR

DIAGNOSTICS

PREV NEXT

16

I/O SETUP COMPLETE

PRESS

ENTER TO CONTINUE

DIAGNOSTICS

PREV

UPDATING EEPROM

DIAGNOSTICS

UPDATING EEPROM

PRESS ENTER

TO RETURN TO

MAIN MENU

DIAGNOSTICS

Action

Select the type of solution for which temperature compensation should be applied.

If the hardware supports analog outputs, specify for each which process variable or computed value will be associated with the output. 3

If analog outputs will be used, specify the scaling for each output. 4

CAT (Current Adjusting Type) control uses Fixed Linear Range only. Retransmitted process values can be scaled. A retransmitted computed value always uses Fixed Linear Range.

At this point, you have completed the I/O setup menu. If you have changed at least one parameter (other than solution temperature compensation) the unit will perform a cold reset (see page 5-1).

When you press ENTER to continue, the message “UPDATING EEPROM” will provide feedback, letting you know your selections have been activated.

When the update has been completed, you will be prompted to exit the menu.

3If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show “DE ENABLED”, instead of displaying output assignments.

4This output range scaling has no effect on the display. Display resolution is configured using the maintenance menu as described in Section 10.

7/99

9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual

5-9