I/O Setup and System Configuration
Step | Screen |
|
|
13SOLUTION TEMP COMP
C1 | ACID/NH3/ETA |
C2 | NACL |
DIAGNOSTICS
PREV NEXT
14ANALOG OUTPUTS
OUTPUT1 C1
OUTPUT2 C2
OUTPUT3 RATO
DIAGNOSTICS
PREV NEXT
15OUTPUT SCALING OUTPUT1 LINFIXEDRNG
OUTPUT2 LINAUTORNG
OUTPUT3
DIAGNOSTICS
PREV NEXT
16
I/O SETUP COMPLETE
PRESS
ENTER TO CONTINUE
DIAGNOSTICS
PREV
UPDATING EEPROM
DIAGNOSTICS
UPDATING EEPROM
PRESS ENTER
TO RETURN TO
MAIN MENU
DIAGNOSTICS
Action
Select the type of solution for which temperature compensation should be applied.
If the hardware supports analog outputs, specify for each which process variable or computed value will be associated with the output. 3
If analog outputs will be used, specify the scaling for each output. 4
CAT (Current Adjusting Type) control uses Fixed Linear Range only. Retransmitted process values can be scaled. A retransmitted computed value always uses Fixed Linear Range.
At this point, you have completed the I/O setup menu. If you have changed at least one parameter (other than solution temperature compensation) the unit will perform a cold reset (see page
When you press ENTER to continue, the message “UPDATING EEPROM” will provide feedback, letting you know your selections have been activated.
When the update has been completed, you will be prompted to exit the menu.
3If the system uses a DE option card, no analog outputs are supported by the hardware. The screen will show “DE ENABLED”, instead of displaying output assignments.
4This output range scaling has no effect on the display. Display resolution is configured using the maintenance menu as described in Section 10.