Conductivity/Resistivity Analyzer/Controller
Watertight corrosion-resistant case
The 9782 is enclosed in a watertight and corrosion resistant industrial case, designed for panel, pipe or wall mounting. It has an EMI/RFI shielded plastic case.
Solution Temperature Compensation
There are two types of 9782C Analyzer/Controller models: standard range and wide range. Both types provide temperature compensation as described below. The standard range Analyzer/Controller is ideal for
To take advantage of the temperature compensation capabilities, the appropriate solution temperature compensation type must be entered for each cell as described in Section 5.
Standard Range - A standard range 9782 Analyzer/Controller provides true high purity temperature compensation for deionized water as well as for routine applications. The Analyzer/Controller also employs unique compensation algorithms for the highest accuracy in power industry water treatment using ammonia or morpholine, as well as for cation conductivity samples. Semiconductor resistivity measurements on acid etch rinse operations, especially with hot deionized water, also achieve the highest available accuracy. For pharmaceutical applications, raw conductivity data is available at the touch of a button.
For
For power plant cation/ammonia conductivity measurements temperature compensation provides the highest accuracy for measurements of cation exchanger effluent and acidic cation conductivity samples with Larson
This compensation matches both strong inorganic and weak organic acid temperature characteristics. Measurements of steam and condensate samples containing ammonia are also greatly improved using this algorithm.
For power plant morphonline conductivity measurements temperature compensation was developed specifically for this market’s water chemistry. Morpholine has unique temperature characteristics that are handled automatically by the 9782 Analyzer/Controller.