I/O Setup and System Configuration
Step | Screen |
|
|
3CAT CONTROL
CAT1
CAT2
CAT3
DIAGNOSTICS
PREV NEXT
4CAT CONTROL OUTPUT 1 SCALING LINFIXEDRNG
INPUTC1
SETPOINT | 12000μS | |
PB LIMIT | 300μS | |
|
|
|
DIAGNOSTICS |
|
|
PREV | NEXT |
|
5RETRANSMISSION OUTPUT 1
OUTPUT 2
DIAGNOSTICS
PREV NEXT
6OUTPUT 1
SCALING LINFIXEDRNG
| INPUT | C1 |
|
| 0% | 0.0µS |
|
| 100% | 19.0µS |
|
|
|
|
|
| DIAGNOSTICS |
|
|
| PREV | NEXT |
|
|
|
|
|
Action
Select an output to configure it for control.
Go to the next screen.
The input type and appropriate unit of measure will already be filled in, based on the process variable or computed value assigned to this analog output during I/O setup. The scaling characteristic will also be filled in; CAT output is always linear
Use the setpoint and proportional band parameters to define the output span.
To configure retransmission on a unit with a single analog output, go to CONFIGURATION
CAT/RETRANSMISSION RETRANSMISSION.
To configure retransmission on a unit with three analog outputs go to CONFIGURATION CAT/RETRANSMISSION and select the output to be configured.
For each output to which a process variable or computed value was assigned during I/O setup, the value assigned will be displayed, as well as the type of scaling to be applied, and the appropriate unit of measure. (This is
Specify the range by entering the values to be represented by outputs of 0% and 100%. If
7/99 | 9782 Series Conductivity/Resistivity Analyzer/Controller - Operator’s Manual |