Chapter 3 Service Adjustments
3.2Arc Lamp Current Setting
The Arc Lamp current must be set whenever an Arc Lamp or Arc Lamp Power Supply is replaced. Routinely recheck current about every 500 hours of operation. This procedure is performed using two DVMs, for convenience, but can be performed with one DVM by moving the leads back and forth between the two test points mentioned and changing the scale ranges on the DVM.
To perform the Arc Lamp Current Setting:
1.Verify that power is off to the projector, or turn power off with the remote (it is not necessary for the circuit breaker to be off).
2.Place a DVM (set to read up to 200VDC) across the Arc Lamp DC INPUT ± terminals, J74, (Figure
3.Turn projector power on.
Figure |
4.Place another DVM, set to read at least 200 millivolts dc, across the test points inside the small rectangular aperture, near the upper left of the Arc Lamp Power Supply cover (see Figure
Model 200 Service Manual |