![](/images/new-backgrounds/1191448/191448169x1.webp)
Chapter 3 Service Adjustments
4.Select the Focus test pattern.
5.Select Projection Lens from the menu.
6.Use the up/down arrow keys to zoom the Projection Lens to full telephoto position (smallest image). NOTE: Skip this step when using an optional fixed lens.
7.Use left and right arrow keys to focus the projection lens to get sharply focused spacer balls. NOTE 1: Spacer balls are used inside the ILA® Assembly to separate the layers. They are tiny, random,
8.Zoom the lens to a wide angle position (largest image). NOTE: Skip this step when using an optional fixed lens.
9.Put on safety gloves (see Safety section for gloves type) then loosen the wing nut on the Green Relay Lens (or whichever lens is being focused). NOTE: If the ILA®/Relay Lens/CRT assembly cannot be moved close enough to get a good spacer ball focus, loosen the CRT Lens Stopper Ring in front of the CRT Holder Ring (Figure
10.Loosen the 2 hex bolts (Figure
11.Slide the Relay Lens/CRT/ILA® assembly forward or backward until the sharpest spacer ball focus is achieved. NOTE: Do not use the zoom lens focus while performing this step.
12.Repeat Steps
13.Tighten the hex bolts on the Green ILA® Assembly and the wing nut on the Green Relay Lens.
14.Repeat Steps
NOTE: The CRT Mechanical Focus procedure (Section 3.5) is often performed sequentially with the CRT Rotation procedure (Section 3.6). The same two set screws (Figure
Model 200 Service Manual |