CESE (Combined Event Status Enable)
Command :CESE <value>
The CESE command sets the Combined Event Status Enable register. This
register is the enable register for the CESR register and contains the
combined status of all of the MESE (Module Event Status Enable) registers
of the Agilent 1670G-series logic analyzers. Table 9-3 lists the bit values for
the CESE register.
<value> An integer from 0 to 65535
Example OUTPUT XXX;":CESE 32"
Query :CESE?
The CESE? query returns the current setting.
Returned Format [:CESE] <value><NL >
Example OUTPUT XXX;":CESE?"
Table 9-3 Agilent 1670G-Series Combined Event Status Enable Register
Bit Weight Enables
3 to 15 not used
2 not used
1 2 logic analyzer
0 1 Intermodule
Instrument Commands
CESE (Combined Event Status Enable)
9–9