NAME
Command :MACHine{1|2}:NAME <machine_name>
The NAME command allows you to assign a name of up to 10 characters to a
particular analyzer (machine) for easier identification.
<machine_name> string of up to 10 alphanumeric characters
Example OUTPUT XXX;":MACHINE1:NAME ’DRAMTES T’"
Query :MACHine{1|2}:NAME?
The NAME query returns the current analyzer name as an ASCII string.
Returned Format [:MACHine{1|2}:NAME] <machine name><NL>
Example OUTPUT XXX;":MACHINE1:NAME?"
REName
Command :MACHine{1|2} :REName {{<res _i d>, <new_text> } |

DEFault}

The REName command allows you to assign a specific name of up to eight
characters to terms A through J, Range 1 and 2, and Timer 1 and 2 in the
state analyzer. In the timing analyzer, EDGE 1 and 2 can be renamed in
addition to the terms available in the state analyzer minus H and J. The
DEFault option sets all resource term names to the default names assigned
when turning on the instrument.
<res_id> {<state_terms>|H|J} for state analyzer
{<state_terms>|EDGE{1|2}} for timing analyzer
MACHine Subsystem
NAME