FALLtime?
Query :MEASure:[SOU Rce CHANnel<N> ;] FALLtime?
The FALLtime query makes a fall time measurement on the selected channel.
The measurement is made between the 90% to the 10% voltage point of the
first falling edge displayed on screen. If a parameter cannot be measured, the
instrument responds with 9.9E37.
Returned Format [:MEASure:FALLti me ] <value><NL>
<N> {1|2}
<value> time in seconds between the 90% and 10% voltage points of the first falling
edge displayed on the screen
Example OUTPUT XXX;":MEASURE:SOUR CHAN2;FALLTIME?"
FREQuency?
Query :MEASure:[SOU Rce CHANnel<N> ;] FREQuency?
The FREQency query makes a frequency measurement on the selected
channel. The measurement is made using the first complete displayed cycle
at the 50% voltage level. If a parameter cannot be measured, the instrument
responds with 9.9E37.
Returned Format [:MEASure:FREQuency]<value><NL>
<N> {1|2}
<value> frequency in Hertz
Example OUTPUT XXX;":MEASUR E:SOUR CHAN1;FREQ? "
MEASure Subsystem