VAMPlitude?
Query :MEASure:[SOU Rce CHANnel<N> ;] VAMPlitude?
The VAMPlitude query makes a voltage measurement on the selected
channel. The measurement is made by finding the relative maximum (VTOP)
and minimum (VBASe) points on screen. If a parameter cannot be
measured, the instrument responds with 9.9E37.
Returned Format [:MEASure:VAMPli tu de] <value><NL>
<N> {1|2}
<value> difference between top and base voltage
Example OUTPUT XXX;":MEASURE:SOURCE CHANNEL2;VAMP?"
VBASe?
Query :MEASure:[SOU Rce CHANnel<N> ;] VBASe?
The VBASe query returns the base voltage (relative minimum) of a displayed
waveform. The measurement is made on the selected source. If a parameter
cannot be measured, the instrument responds with 9.9E37.
Returned Format [:MEASure:VBASe] <value><NL>
<N> {1|2}
<value> voltage at base (relative minimum) of selected waveform
Example OUTPUT XXX;":MEASURE:SOURCE CHAN1;VBAS?"
MEASure Subsystem
VAMPlitude?
33-10