Table 9-6 Agilent 1670G-Series Logic Analyzer Mainframe (Intermodule) Module Event Status
Enable Register
Bit Position Bit Weight Enables
7 128 not used
6 84 not used
5 32 not used
4 16 not used
3 8 not used
2 4 not used
1 2 RNT - Intermodule Run Until Satisfied
0 1 MC - Intermodule Measurement Complete
Table 9-7 Agilent 1670G-Series Logic Analyzer Module Event Status Enable Register
Bit Position Bit Weight Enables
7 128 not used
6 84 not used
5 32 not used
4 16 not used
3 8 Pattern searches failed
2 4 Trigger found
1 2 RNT - Run Until Satisfied
0 1 MC - Measurement Complete
Instrument Commands
MESE<N> (Module Event Status Enable)