Tag Data Description
If tags are enabled for one or both analyzers, the tag data follows the
acquisition data. The first byte of the tag data is determined as follows:
591 + (bytes per row × maximum number of valid rows)
Each row of the tag data array consists of one (single tags enabled) or two
(both analyzer’s tags enabled) eight-byte tag values per row. When both
analyzers have tags enabled, the first tag value in a row belongs to analyzer
number one and the second tag value belongs to analyzer number two.
If the tag value is a time tag, the number is an integer representing time in
picoseconds. If the tag value is a state tag, the number is an integer state
count.
The total size of the tag array is eight or 16 bytes per row (as described in
Acquisition Data Description on page 27-10) times the greatest number of
valid rows.
SYSTem:SETup
Command :SYStem:SETup <block data>
The SYStem:SETup command configures the logic analyzer module as
defined by the block data sent by the controller.
Three data sections are always included. These are the strings which would
be included in the section header.
"CONFIG "
"DISPLAY1 "
"BIG_ATTRIB"
Additionally, the following sections may also be included, depending on
what’s available:
"SYMBOLS A "
"SYMBOLS B "
"INVASM A "
"INVASM B "
DATA and SETup Commands
Tag Data Description
27–12