MODe
The MODe command is used to specify either FULL or HALF channel output
mode. Half channel mode allows a higher output data rate (greater than 100
MHz), but with only 20 channels per .
Full channel output mode limits the maximum data rate to 100 MHz but
allows use of 40 channels per .
The output mode selection sets the upper limit for the clock rate (see
FORMat:CLOCk command).
Command syntax: :FORMat:MODe{FULL|HALF}
Query syntax: :FORMat:MODe?
Returned format: [FORMat:MODe]{FULL|HALF}
Assigning labels in half-channel mode erases previously-assigned labels.
FORMat Subsystem
MODe
38–7