Terms A through E, RANGE 1, and EDGE1 must be grouped together and terms
F, G, RANGE 2, and EDGE2, and must be grouped together. In the first level,
terms from one group may not be mixed with terms from the other. For example,
the expression ((A OR IN_RANGE2) AND (C OR G)) is not allowed
because the term C cannot be specified in the F, G group.
In the first level, the operators you can use are AND, NAND, OR, NOR,
XOR, NXOR. Either AND or OR may be used at the second level to join the
two groups together. It is acceptable for a group to consist of a single term.
Thus, an expression like (B AND G) is legal since the two operands are both
simple terms from separate groups.
CLEar
Command :MACHine{1|2}:TTRigger:CLEar
{All|SEQuence|RESource}
The CLEar command allows you to clear all settings in the Timing Trigger
menu and replace them with the default, clear only the sequence levels, or
clear only the resource term patterns.
Example OUTPUT XXX;":MACHINE1:TTRIGGER:CLEAR RESOURCE"
TTRigger (TTRace) Subsystem
CLEar
22–12