ACQMode
Command :MACHine{1|2}:TFORmat:AC QMode {FULL | HALF}
The ACQMode (acquisition mode) command selects the acquisition mode for
the timing analyzer. The options are:
conventional mode at full-channel 125 MHz
conventional mode at half-channel 250 MHz
Example OUTPUT XXX;":MACH INE2:TFORMAT:ACQ MODE HALF"
Query :MACHine{1|2}:TFORmat:ACQMode?
The ACQMode query returns the current acquisition mode.
Returned Format [:MACHine{1|2}:T FO Rmat:ACQMode] {FULL | HA LF}<NL>
Example OUTPUT XXX;":MACHIN E2:TFORMAT:ACQMO DE?"
TFORmat Subsystem
ACQMode