OAUTo
Command :MARKer:OAUTo {MANual|CHAN nel<N>,<type>, <level>,

<slope>,<occurrence>}

The OAUTo command specifies the automatic placement specification for
the O marker. The first parameter specifies if automarker placement is to be
in the manual mode or on a specified channel. If a channel is specified, four
other parameters must be included in the command syntax. The four
parameters are marker type, level, the slope, and the occurrence count.
<N> {1|2}
<type> {ABSolute | PERCent}
<level> percentage of waveform voltage level, ranging from 10 to 90 of the Vtop to
Vbase voltage or a voltage level
<slope> {POSitive | NEGative}
<occurrence> integer from 1 to 100
Example OUTPUT XXX;":MARKER:OAUTO CHANNEL1,PERCent,50,POSITIVE,5"
Query :MARKer:OAUTo?
The OAUTo query returns the current settings.
Returned Format [:MARKer:OAUTo] (M AN ual|CHANnel<N>,<ty pe>
<level>,<slope>,<occurrence>}<NL>
Example OUTPUT XXX;":MARKER :OAUTO?"
MARKer Subsystem
OAUTo
32-9