PWIDth?
Query :MEASure:[SOU Rce CHANnel<N> ;] PWIDth?
The PWIDth query makes a positive pulse width measurement on the
selected channel. The measurement is made by finding the time difference
between the 50% points of the first rising and the next falling edge displayed
on screen. If a parameter cannot be measured, the instrument responds with
9.9E37.
Returned Format [:MEASure:PWIDth ] <v alue><NL>
<N> {1|2}
<value> positive pulse width in seconds
Example OUTPUT XXX;":MEASURE:SOURCE CHANNEL2;PWIDTH?"
RISetime?
Query :MEASure:[SOU Rce CHANnel<N> ;] RISetime?
The RISetime query makes a risetime measurement on the selected channel
by finding the 10% and 90% voltage levels of the first rising edge displayed on
screen. If a parameter cannot be measured, the instrument responds with
9.9E37.
Returned Format [:MEASure:RISeti me ] <value><NL>
<N> {1|2}
<value> risetime in seconds
Example OUTPUT XXX;":MEASU RE:SOUR CHAN1;RI SETIME?"
MEASure Subsystem
PWIDth?
33-8