MESE<N> (Module Event Status Enable)
Command :MESE<N> <enabl e_value>
The Agilent 1670G-series logic analyzers support the MESE command for
compatibility with other logic analyzer programs but do not take any action
when the command is sent. In 16500 programs, the MESE command sets the
Module Event Status Enable register. The <N> index specifies the module,
and the parameter specifies the enable value.
<N> An integer, 0 through 10.
<enable_value> An integer from 0 through 255.
Example OUTPUT XXX;":MESE1 3"
Query :MESE<N>?
The query returns the current setting. Tables 9-6 and 9-7 list the Module
Event Status Enable register bits, bit weights, and what each bit masks for
the mainframe and logic analyzer respectively.
Returned Format [:MESE<N>] <enabl e_value><NL>
Example OUTPUT XXX;":MESE1?"
Instrument Commands
MESE<N> (Module Event Status Enable)
9–14