The print query should NOT be sent with any other command or query on the
same command line. The print query never returns a header. Also, since
response data from a print query may be sent directly to a printer without
modification, the data is not returned in block mode.
Example OUTPUT 707;":SYSTEM:PRINT? SCREEN"
SETupCommand :SYStem:SETup <block_data>
The :SYStem:SETup command configures the logic analyzer module as
defined by the block data sent by the controller. This chapter describes
briefly the syntax of the Setup command and query. Because of the
capabilities and importance of the Setup command and query, a complete
chapter is dedicated to it. The dedicated chapter is chapter 27, "DATA and
SETup Commands."
<block_data> <block_length_specifier><section>
<block_length_
specifier> #8<length>
<length> The total length of all sections in byte format (must be represented with 8
digits)
<section> <section_header><section_data>
<section_
header> 16 bytes, described in the "Section Header Description" section in chapter 27.
<section_data> Format depends on the type of data
The total length of a section is 16 (for the section header) plus the length of
the section data. When calculating the value for <length>, don’t forget to
include the length of the section headers.
Example OUTPUT XXX USING "#,K";":SYSTEM:S ETUP " <block_d ata>
SYSTem Subsystem
SETup