Table 31-1 DISPlay Parameter Values
Parameter Value
slot_# 1 or 2 1=analyzer, 2=oscilloscope.
bit_id an integer from 0 to 31.
channel_# 1 or 2.
label_str up to five characters enclosed in single quotes making up a
label name.
label_id a string of 1 alpha and 1 numeric character for the
oscilloscope, or 6 characters for the timing modules.
ACCumulate
Command :DISPlay:ACCumulate {{ON|1 }|{OFF|0}}
The ACCumulate command works in conjunction with the commands in the
Acquisition Subsystem. In the Normal mode, the ACCumulate command
turns infinite persistence on or off.
Example OUTPUT XXX;":DISPLAY:ACC ON"
Query :DISPLAY:ACCumulate?
The ACCumulate query reports if accumulate is turned on or off.
Returned Format [:DISPlay:ACCumu la te] {1|0}<NL>
Example OUTPUT XXX;":DISPLAY:ACCUMULATE?"
DISPlay Subsystem
ACCumulate
31-4