End points for the INRange and OUTRange should be at least 8 ns apart since
this is the minimum time resolution of the time tag counter.
There are two conditions which are based on a comparison of the acquired
state data and the compare data image. You can run until one of the
following conditions is true:
Every channel of every label has the same value (EQUal).
Any channel of any label has a different value (NEQual).
The RUNTil instruction (for state analysis) is available in both the SLISt and
COMPare subsystems.
<value> real number from 9E9 to +9E9
Example OUTPUT XXX;":MACHIN E2:COMPARE:RUNTI L EQUAL"
Query :MACHine{1|2}:COMPare:RUNTil?
The RUNTil query returns the current stop criteria for the comparison when
running in repetitive trace mode.
Returned Format [:MACHine{1|2}:COMPare:RUNTil] {OFF| LT,<value>|GT,<value>l
INRange,<value>,<value>|OUTRange,<value>,<value>|EQUal|NEQual}
<NL>
Example OUTPUT XXX;":MACHIN E2:COMPARE:RUNTI L?"
COMPare Subsystem
RUNTil (Run Until)
20–11