APC AXFA11G Standard Performance, Reference Conditions, Pulse Output PFA/Ceramics Lining, Outline

Models: AXFA11G

1 118
Download 118 pages 18.28 Kb
Page 111
Image 111
■STANDARD PERFORMANCE

STANDARD PERFORMANCE

Reference Conditions:

Similar to BS EN 29104 (1993); ISO9104 (1991)

Fluid Temperature: 20°C ±10°C (+68°F ±18°F)

Ambient Temperature: 25°C ±5°C (+77°F ±9°F)

Warm-up Time: 30 min

Straight runs

Upstream > 10  DN

Downstream > 5  DN

Properly grounded

Properly centered

Accuracy (Combined with AXF Remote Flowtube, at reference conditions)

Pulse Output:

PFA/Ceramics Lining:

Size mm

Flow Velocity

Standard

Flow Velocity

 

High grade

Accuracy

 

Accuracy

(in.)

V m/s (ft/s)

(Calibration

V m/s (ft/s)

 

(Calibration

 

 

 

code B)

 

 

 

code C)

 

 

 

 

 

 

 

2.5 (0.1)

V 0.3 (1)

1.0 mm/s

 

 

 

 

 

 

 

 

 

to

0.3  V  10

0.35% of

 

 

15 (0.5)

 

 

 

 

(1)

(33)

Rate

 

 

 

 

 

 

 

 

 

 

 

V 0.15 (0.5)

0.5 mm/s

V 0.15 (0.5)

 

0.5 mm/s

25 (1.0)

 

 

 

 

 

 

 

 

 

 

0.15  V 1

 

0.18% of Rate

to

0.15  V  10

0.35% of

 

(0.5)

(3.3)

 

0.2mm/s

200 (8.0)

(0.5)

(33)

Rate

 

 

 

 

1  V 10

0.2% of Rate

 

 

 

 

 

 

 

 

(3.3)

(33)

 

 

 

 

 

 

 

 

 

250 (10)

V 0.15 (0.5)

0.5 mm/s

 

 

 

 

 

 

 

 

 

to

0.15  V  10

0.35% of

 

 

400 (16)

(0.5)

(33)

Rate

 

 

 

 

 

 

 

 

 

 

 

 

Enhanced dual frequency excitation (Option code HF2) :

T01.EPS

 

Standard accuracy  1 mm/ s

 

Size 2.5 mm (0.1 in.) to 15 mm (0.5 in.)

%of Rate 1.2

 

1.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pulse

0.8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HighStandardgr eAccuracyA curacy

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accuracy 0.4

 

 

 

 

 

 

 

 

 

0.35

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0 1 26 8 10 V[m/s]

F01.EPS

Size 25 mm (1.0 in.) to 400 mm (16 in.)

%of Rate 1.2

 

1.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pulse

0.6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Standard Accuracy

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

High grade Accuracy

 

Accuracy

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.35

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0 1 26 8 10 V[m/s]

F02.EPS

11. OUTLINE

Polyurethane Rubber/Natural Soft Rubber/EPDM Rubber Lining

Size mm (in.)

Flow Velocity

Standard Accuracy

V m/s (ft/s)

(Calibration code B)

 

 

 

 

25 (1.0)

V 0.3 (1.0)

1.0 mm/s

to

0.3  V  10

0.35% of Rate

400 (16)

(1.0)

(33)

 

 

 

 

 

 

V 0.3 (1.0)

1.75 mm/s

500 (20)

 

 

0.3  V 1

0.25% of Rate 1 mm/s

to

(1.0)

(3.3)

1000 (40)

 

 

 

1  V

 10

0.35% of Rate

 

 

(3.3)

(33)

 

 

 

 

V 0.3 (1.0)

2.2 mm/s

 

 

 

1100 (44)

0.3  V 1

0.4% of Rate 1 mm/s

to

(1.0)

(3.3)

 

2000 (80)

 

 

 

1  V

 10

0.5% of Rate

 

 

(3.3)

(33)

 

 

 

2200 (88)

V 1 (3.3)

8.5 mm/s

to

 

 

 

1  V

 10

0.85% of Rate

2600 (104)

(3.3)

(33)

 

 

 

 

 

 

Enhanced dual frequency excitation (Option code HF2) : T02.EPS Standard accuracy  1 mm/ s

Size 25 mm (1.0 in.) to 400 mm (16 in.)

%of Rate 1.2

 

1.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pulse

0.8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HStandardigh gr eAccuracyA curacy

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accuracy 0.4

 

 

 

 

 

 

 

 

 

0.35

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0 1 26 8 10 V[m/s]

F01.EPS

Size 500 mm (20 in.) to 1000 mm (40 in.)

%of Rate 1.2

 

1.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pulse

0.8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HStandardigh gr eAccuracyA curacy

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Output

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Accuracy 0.4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.35

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0.2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

1

 

 

 

 

 

 

2

6

8

10

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

V[m/s]

 

F16-1.EPS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Size 1100 mm (44 in.) to 2000 mm (80 in.)

%of Rate 1.2

1.0

 

0.8

HStandardigh gr eAccuracyA curacy

 

Pulse 0.6

Output 0.5

Accuracy 0.4

0.2

0

0 1 26 8 10 V[m/s]

F16-2.EPS

11-3

IM 01E20C01-01E

Page 111
Image 111
APC AXFA11G user manual Standard Performance, Accuracy Combined with AXF Remote Flowtube, at reference conditions, Outline