8. OPERATION VIA HART COMMUNICATOR
NOTE
Data changed with the HART Communicator is sent to the AXFA11 by pressing
SEND (F2) of the HART Communicator.
8.3.3Checking for Problems
The
* Open the “Test/Status” setting display.
1.Device Setup ![]()
![]()
2. Diag/Service ![]()
![]()
3. Test/Status
8.3.4Setting Parameters
As mentioned in Section 5.2.2 “Setting Mode,” the AXFA11’s parameters are divided into three types:
i.Selection type
ii.Numeric type
iii.Alphanumeric type
This section describes how to set these parameters using a HART communicator.
NOTE
All three parameters must be set to obtain a correct signal.
Nominal size, flow rate span and meter factor must be set.
8.3.4.1Example of Setting Selection-Type Data: Base Flow Unit
This example describes the setting of the flow rate units for the
Open the “Base Flow Unit” setting display.
1. Device Setup ![]()
![]()
3. Easy Setup ![]()
![]()
3. Base Flow Unit
1AXFA11 :
m3
m3
K1(Kiloliter)
l(Liter)
cm3
ESC ENTER
2AXFA11 :
1 Language
2 | PV | Damping |
| |
3 | Base | Flow Unit | l | |
4 | Base | Time Unit | /h | |
5 | PV | Span |
| |
SEND HOME
⋅Select “1(Liter)”.
2
F4 | Press ENTER (F4). |
|
|
(ENTER)
3AXFA11 :
Easy Setup
1Language
2PV Damp
F2
Press SEND (F2) to set the data into the AXFA11.
3 | Base | Flow Unit | l |
4 | Base | Time Unit | /h |
5 | PV Span |
| |
SAVE HOME
(SEND)
F0807.EPS