6. PARAMETER DESCRIPTION

Parameter setting sequence

Output

 

 

(for automatic multiple ranges switching)

No. 1 range

No. 2 range No. 3 range

No. 4 range

 

 

 

 

100%

F10: SO1 Function

F11: SO2 Function

 

 

 

Select a function

Select a function

 

 

 

F14: SO1/2 Active Mode

 

 

 

Select whether SO1 and SO2 output is to

 

 

 

be “Closed (On) Act” or “Open (Off) Act”

 

 

 

 

 

 

 

Input

Function Selection

 

 

 

 

No Function: Output is stopped (*: other functions may be selected)

 

10% of No. 3 range

Fwd/Rev Ranges: Forward and reverse flow rate measurement

 

10% of No. 2 range

(set using F40)

Auto 2 Ranges: Automatic 2-ranges switching

 

 

(set using F40)

Auto 3 Ranges: Automatic 3-ranges switching

 

10% of No. 1 range

 

Auto 4 Ranges: Automatic 4-ranges switching

 

(set using F40)

 

Refer to Table 6.4.1 and set F10 or F11.

B23: Flow Span

 

F30: Forward

 

F31: Forward

 

 

Span 2

 

Span 3

Forward No. 1 range

 

Forward No. 2 range

 

Forward No. 3 range

 

 

 

 

 

 

 

 

 

F32: Forward

 

F33: Reverse

 

F34: Reverse

Span 4

 

Span 1

 

Span 2

Forward No. 4 range

 

Reverse No. 1 range

 

Reverse No. 2 range

 

 

 

 

 

Set the spans for the ranges to be used.

No. 1 range  No. 2 range

No. 3 range  No. 4 range

If “instantaneous flow rate % display (FR)” is selected for

 

Display Mode and A10: FLOW RATE(%) is selected for

 

BRAIN communication, the instantaneous flow rate %

 

and following symbols will be displayed only for multiple

 

ranges and forward/reverse flow measurement.

 

Forward No. 1 range : [F1]

Reverse No. 1 range : [R1]

 

Forward No. 2 range : [F2]

Reverse No. 2 range : [R2]

 

Forward No. 3 range : [F3]

 

 

Forward No. 4 range : [F4]

 

 

 

 

 

 

F0607.EPS

Reverse

Forward

Minimum span % within multiple and

 

forward/reverse (set using F41)

F0608.EPS

Figure 6.4.1 Multiple Ranges and Hysteresis Widths

NOTE

For more details regarding the setting of hyster- esis width, refer to the description of setting

parameter for F40: Auto Range Hys and F41: Bi Direction Hys.

Multiple Ranges Setting 2:

Multiple ranges switching via external status input

￿For both the forward and reverse directions, switch- ing of up to four ranges can be carried out based on status input; however, switching between directions is not possible. Switching between forward and reverse ranges is carried out automatically only when the flow direction reverses.

￿SI1 and SI2 status input terminals are used for multiple range switching. For more details, refer to Table 6.4.2: Multiple Ranges Switching via External Status Input.

6-23