Image 64
6.5.2Alarm Selection

6. PARAMETER DESCRIPTION

6.5.2Alarm Selection

The display and output differs depending on the alarm levels. Certain types of alarm may or may not be recognized as alarms, according to the settings of certain parameters. The parameters that are relevant to this function as follows.

[G20: Alm Out Act Mode] Setting of the active

mode for the alarm output

[G21: 4-20mA Alarm Out] Setting of the current output during an alarm occurring.

[G30: Alm-Setting]Alarm recognition of “Setting Alarm”

[G31: Alm-Sig Over] Alarm recognition of “Signal Overflow Alarm”

[G32: Alm-Emp Pipe] Alarm recognition of “Empty Pipe alarm”

[G33: Alm-HH/LL]Alarm recognition of “HH/LL Alarm” (Refer to the descriptions of G12 and G13 for more details regarding HH and LL alarms.)

[G34: Alm-Adhesion]Alarm recognition of “Adhe- sion Alarm”

[G41: Alm Record1] Alarm record1

[G43: Alm Record2] Alarm record2

[G45: Alm Record3] Alarm record3

[G47: Alm Record4] Alarm record4

(1) Display and output condition for system alarms

 

 

Alarm description

Alarm

4-20 mA

Totali-

Pulse

Display unit

Alarm

 

 

output

output

zation

record

 

 

 

 

 

 

 

 

 

 

Normal

 

Closed

Normal

Normal

Normal

Display Mode

No

 

 

(On)

 

 

 

 

 

 

 

 

10

P Fault

Microprocessor(CPU)failure

Open

0mA or

Indeter-

Stopped

Indetermination

Indeter-

11

EEPROM Fault

EEPROM failure

(Off)

25mA (*)

mination

 

 

mination

 

 

 

 

 

 

12

A/D(H) Fault

 

Open

Fixed

 

 

Alarm Mode

 

13

A/D(L) Fault

A/D converter failure

Stopped

Stopped

(display of system

Recorded

(Off)

(G21 selection)

 

 

 

 

 

alarm message)

 

14

A/D(Z) Fault

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

15

Coil Open

Flowtube coil is open-circuit

 

 

 

 

 

 

16

EEPROM Dflt

EEPROM default values

 

 

 

 

 

 

Note: • Operations are performed in accordance with above table, when “Open (Off) Act” is set for G20: Alm

Out Act Mode.

4-20 mA output upon the occurrence of an alarm will be fixed at the value selected with G21: 4-20mA Alarm Out.

*The output value is performed in accordance with the setting of the burnout switch. For information about

this switch, see Section 10.2.1.T0648.EPS

(2) Display and output condition for process alarms

 

 

Alarm

Selection

Alarm

4-20 mA

Totali-

Pulse

Display

Alarm

 

 

(parameter

 

 

description

output

output

zation

output

unit

record

 

 

 

number)

 

 

 

 

 

 

 

 

 

YES

Open

Fixed

Stopped

Stopped

Alarm Mode

Recorded

30

Sig Overflow

Input signal error

(G31)

(Off)

 

 

 

(Message)

 

NO

Closed

Continu-

Continu-

Continu-

Display Mode

No

 

 

 

 

 

 

(G31)

(On)

ous (*)

ous (*)

ous (*)

 

 

Flowtube is not

YES

Open

Fixed

Stopped

Stopped

Alarm Mode

Recorded

31

Empty Pipe

(G32)

(Off)

 

 

 

(Message)

 

filled with fluid

NO

Closed

Continu-

Continu-

Continu-

Display Mode

No

 

 

 

 

 

 

 

 

(G32)

(On)

ous (*)

ous (*)

ous (*)

 

 

 

 

 

YES

Open

 

 

 

Alarm Mode

 

32

HH/LL Alm

HH/LL Alarm

(G33)

(Off)

Normal

Normal

Normal

(Message)

No

NO

Closed

operation

operation

operation

Display Mode

 

 

 

 

 

 

 

(G33)

(On)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Electrode

YES

Open

Fixed

Stopped

Stopped

Alarm Mode

Recorded

33

Adhesion Alm

(G34)

(Off)

 

 

 

(Message)

 

adhesion alarm

NO

Closed

Continu-

Continu-

Continu-

Display Mode

No

 

 

 

(G34)

(On)

ous (*)

ous (*)

ous (*)

Note: • Operations are performed in accordance with above table, when “Open (Off) Act” is set for G20: Alm Out Act Mode.

4-20 mA output upon the occurrence of an alarm will be fixed at the value selected with G21: 4-20mA Alarm Out.

T0649.EPS

*: Although outputs are continuous, output values are not guaranteed.

6-36

IM 01E20C01-01E