6. PARAMETER DESCRIPTION
Setting Alarms (Device is normal but errors have been made in the setting of parameters.) | ||||
Display unit/BRAIN | Alarm countermeasure | Alarm description | Countermeasure | |
(60) content | message on display unit | |||
|
| |||
|
| Span flow velocity setting is 11 m/s or | Check whether parameters | |
50:Span > 10m/s | Check parameter C40, C41, | more | C40, C41, and C42 are correct. | |
| In case that multiple range or | |||
and C42 |
| |||
|
| forward and reverse flow | ||
|
|
| ||
|
|
| measurement functions is used, | |
51:Span < 0.1m/s | Check parameter C40, C41, | Span flow velocity setting is 0.05 m/s or | check whether parameters F30 | |
and C42 | less | through F36 are correct. | ||
| ||||
52:TTL>10000p/s | Check parameter D10 and | Totalization rate is 11000 pps or more |
| |
D11 |
| Check whether parameters | ||
|
| |||
53:TTL<0.0001p/s | Check parameter D10 and | Totalization rate is 0.00005 pps or less | D10 and D11 are correct. | |
D11 |
|
| ||
|
|
| ||
Check parameter J11and J12 | The condition | Check whether parameters J11 | ||
|
| and J12 are correct. | ||
55:Multi Rng Err |
| The condition [No. 1 range < No. 2 | Check whether parameters | |
Check parameter F30 to F36 | range < No. 3 range < No. 4 range] is | |||
F30 through F36 are correct. | ||||
|
| not satisfied for multiple ranges. | ||
|
|
| ||
|
|
|
| |
|
| The condition [High Alarm (G11) – |
| |
56:H/L HH/LL Set |
| Low Alarm (G10) > H/L Alarm Hys | Check whether parameters | |
Check parameter G10 to G14 | (G14)] or the condition [High High | |||
G10 through G14 are correct. | ||||
|
| Alarm (G13) – Low Low Alarm (G12) | ||
|
|
| ||
|
| > H/L Alarm Hys (G14)] is not satisfied. |
| |
| Check parameter C40, C45, | Mass units have been selected for Base | Check whether parameters | |
57:Dens Set Err | Flow Unit (C40) but density is set to 0. | |||
and C46 | C40, C45, and C46 are correct. | |||
|
| |||
|
|
|
| |
|
| Pulse rate is 11000 pps or more with |
| |
60:PLS > 10000p/s | Check parameter E10, E11, | 50% duty selection. |
| |
and E12 | Pulse rate is 10000 pps or more with |
| ||
|
| |||
|
| 0.05 ms selection. |
| |
61:PLS > 5000p/s | Check parameter E10, E11, | Pulse rate is 5000 pps or more with 0.1 |
| |
and E12 | ms selection. |
| ||
|
| |||
|
|
|
| |
62:PLS > 1000p/s | Check parameter E10, E11, | Pulse rate is 1000 pps or more with 0.5 |
| |
and E12 | ms selection. |
| ||
|
| |||
63:PLS > 500p/s | Check parameter E10, E11, | Pulse rate is 500 pps or more with 1 ms |
| |
and E12 | selection. | Check whether parameters | ||
| ||||
64:PLS > 25p/s | Check parameter E10, E11, | Pulse rate is 25 pps or more with 20 ms | E10, E11, and E12 are correct. | |
and E12 | selection. |
| ||
|
| |||
|
|
|
| |
65:PLS > 15p/s | Check parameter E10, E11, | Pulse rate is 15 pps or more with 33 ms |
| |
and E12 | selection. |
| ||
|
| |||
|
|
|
| |
66:PLS > 10p/s | Check parameter E10, E11, | Pulse rate is 10 pps or more with 50 ms |
| |
and E12 | selection. |
| ||
|
| |||
|
|
|
| |
67:PLS > 5p/s | Check parameter E10, E11, | Pulse rate is 5 pps or more with 100 ms |
| |
and E12 | selection. |
| ||
|
| |||
|
|
|
| |
70:PLS<0.0001p/s | Check parameter E10, E11, | Pulse rate is 0.00005 pps or less. |
| |
and E12 |
| |||
|
|
| ||
|
|
|
| |
71:Meas Mod Set | Check parameter C20 | Measure Mode (C20) is set to Enhanced | Check whether parameter C20 | |
DF without selecting an optional | is correct. | |||
|
| code HF1 or HF2. | ||
|
|
| ||
72: Size Set Err | Check parameter C32 | A value of 3000.1 mm or more is set for | Check whether parameter C32 | |
Nominal Size (C32). | is correct. | |||
|
| |||
73: Adh Set Err | Check parameter K11 to K14 | The condition in Adhesion detection | Check whether parameters K11, | |
| level, Level:1<Level:2<Level:3<Level:4 | K12, K13 and K14 are correct. | ||
|
| is not satisfied. |
| |
|
|
| T0653.EPS |
Setting Alarms (Device and measurements are normal but a warning is issued.) | ||||
Display unit/BRAIN | Alarm countermeasure | Alarm description | Countermeasure | |
(60) content | message on display unit | |||
|
| |||
80:Adhesion Wng | – | Slight adhesion to | Clean and check the electrodes. | |
electrodes. | Refer to parameter K13. | |||
|
| |||
|
|
| Carry out adjustment as follows: | |
|
| Results of automatic zero | • Check if the flowtube is filled with | |
82:Auto Zero Wng | – | fluid. | ||
adjustment are higher than | ||||
• Check if the flow velocity is | ||||
| ||||
|
| the rated values. | ||
|
| completely zero. | ||
|
|
| ||
|
|
| • Check the condition of grounding. | |
|
|
|
| |
|
|
| Confirm whether the flow rate is in | |
83:Fix Cur Wng | – | The current value is fixed. | excess of the upper limit (108%) or | |
below the lower limit | ||||
|
|
| whether upon entry to the Test Mode | |
|
|
| or not. | |
84:Disp Over Wng | – | Overflow in the display digits during | Check whether parameter C43 is | |
(only for display unit) | instantaneous flow rate display. | correct. | ||
| ||||
|
|
|
| |
90:Disp SW Wng |
| Display unit switches are not | Investigate whether the display unit | |
– | cover is fitted or whether the cover’s | |||
(only for display unit) | operating. | |||
glass surface is dirty. | ||||
|
|
| ||
|
|
|
| |
|
|
| T0654.EPS |