*“Joker password” and “Software Seal”

9

AXFA11 : FI-1201 Wrt Protect Menu

1 Write protect Yes

2 Enable Wrt 10min

38.3.4.10Other Functions and OperationsNew Password

4 Software seal Keep

SAVE

 

 

8. OPERATION VIA HART COMMUNICATOR

2

Press

twice, and return to the “Online” menu.

F0822-3.EPS

NOTE

1.Enable Wrt 10min” releases write protection status for 10 minutes. While write protection status is released, it is possible to enter a new password after selecting “New Password”. It will not be possible after the 10 minutes have elapsed.

2.To release write protection status completely, enter 8 spaces in the “New Password”. This causes the “Write protect” status to change from “Yes” to “No”.

3.If the power supply to either the AXFA11 or the HART Communicator are turned off and then on again within 10 minutes after exiting write protection status, the enabled write protection status becomes defunct.

*Joker password” and “Software Seal

When you forget the password that you have input, it is possible to release the mode for 10 minutes by setting a joker password in “Enable Wrt 10 min”. This joker password can be obtained by contacting your Yokogawa sales office.

If this joker password is used, the status shown in the parameter “Software Seal” is changed from “Keep” to “Break”. Press Hot key and select “Wrt Protect Menu”. Current status is shown in “Software Seal”.

This status will be returned from “Break” to “Keep” by registering a new password.

8.3.4.10Other Functions and Operations

Set any other required parameters in the same way as those previous. Refer to the follow- ing document for operation information.

Product Manual for the HART Communicator: MAN 4250

8-20

IM 01E20C01-01E