6. PARAMETER DESCRIPTION

6.5.3Alarms & Warning Messages

System Alarms (Device breakdown or inability to obtain correct measurements.)

Display unit/BRAIN

Alarm countermeasure

Alarm description

Countermeasure

(￿60) content

message on display unit

 

 

10:uP Fault

Contact nearest office or

Microprocessor (CPU) failure

 

service center

 

 

 

 

11:EEPROM Fault

Contact nearest office or

EEPROM failure

 

service center

 

 

 

 

12:A/D(H) Fault

Contact nearest office or

 

Contact your nearest Yokogawa office or

service center

 

service center.

 

 

13:A/D(L) Fault

Contact nearest office or

A/D converter failure

 

service center

 

 

 

 

14:A/D(Z) Fault

Contact nearest office or

 

 

service center

 

 

 

 

 

 

Cut the power and check coil

 

Turn off the AXFA11 power supply and

15:Coil Open

Flowtube coil is open-circuit

examine the excitation cable for breakage.

 

& EX cable

 

If there is no breakage, contact your nearest

 

 

 

Yokogawa office or service center.

16:EEPROM Dflt

Contact nearest office or

EEPROM default values

Contact your nearest Yokogawa office or

service center

service center.

 

 

 

 

 

T0651.EPS

Process Alarms (Device is normal but process-related errors make correct measurement impossible.)

Display unit/BRAIN

Alarm countermeasure

Alarm description

Countermeasure

(￿60) content

message on display unit

 

 

 

 

 

Carry out an investigation as follows:

 

 

 

Check the signal cable for breakage.

30:Sig Overflow

Check signal cable

Input signal error

Check for contact between signal cable,

and grounding

power cable, and excitation cable.

 

 

 

 

 

Check for stray currents in the fluid.

 

 

 

Check the grounding.

 

 

 

 

31:Empty Pipe

Fill flow tube with

Flowtube is not filled with

Fill the flowtube with fluid.

fluid

fluid

 

 

 

 

 

 

 

Check the flow rate and

Flow rate alarm for greater

Check the flow rate and setting value of

32:HH/LL Alm

than High-High limit value

setting value

or less than Low-Low limit

High High limit and Low Low limit.

 

 

 

value.

 

 

 

 

 

33:Adhesion Alm

Clean electrodes

Electrode adhesion alarm

Clean the electrodes.

 

 

 

 

T0652.EPS

6-38

IM 01E20C01-01E

Page 66
Image 66
APC AXFA11G user manual Alarms & Warning Messages