6. PARAMETER DESCRIPTION
6.5.3Alarms & Warning Messages
System Alarms (Device breakdown or inability to obtain correct measurements.) | ||||
Display unit/BRAIN | Alarm countermeasure | Alarm description | Countermeasure | |
(60) content | message on display unit | |||
|
| |||
10:uP Fault | Contact nearest office or | Microprocessor (CPU) failure |
| |
service center |
| |||
|
|
| ||
11:EEPROM Fault | Contact nearest office or | EEPROM failure |
| |
service center |
| |||
|
|
| ||
12:A/D(H) Fault | Contact nearest office or |
| Contact your nearest Yokogawa office or | |
service center |
| service center. | ||
|
| |||
13:A/D(L) Fault | Contact nearest office or | A/D converter failure |
| |
service center |
| |||
|
|
| ||
14:A/D(Z) Fault | Contact nearest office or |
|
| |
service center |
|
| ||
|
|
| ||
| Cut the power and check coil |
| Turn off the AXFA11 power supply and | |
15:Coil Open | Flowtube coil is | examine the excitation cable for breakage. | ||
| & EX cable |
| If there is no breakage, contact your nearest | |
|
|
| Yokogawa office or service center. | |
16:EEPROM Dflt | Contact nearest office or | EEPROM default values | Contact your nearest Yokogawa office or | |
service center | service center. | |||
|
| |||
|
|
| T0651.EPS |
Process Alarms (Device is normal but | ||||
Display unit/BRAIN | Alarm countermeasure | Alarm description | Countermeasure | |
(60) content | message on display unit | |||
|
| |||
|
|
| Carry out an investigation as follows: | |
|
|
| • Check the signal cable for breakage. | |
30:Sig Overflow | Check signal cable | Input signal error | • Check for contact between signal cable, | |
and grounding | power cable, and excitation cable. | |||
|
| |||
|
|
| • Check for stray currents in the fluid. | |
|
|
| • Check the grounding. | |
|
|
|
| |
31:Empty Pipe | Fill flow tube with | Flowtube is not filled with | Fill the flowtube with fluid. | |
fluid | fluid | |||
|
| |||
|
|
|
| |
| Check the flow rate and | Flow rate alarm for greater | Check the flow rate and setting value of | |
32:HH/LL Alm | than | |||
setting value | or less than | High High limit and Low Low limit. | ||
| ||||
|
| value. |
| |
|
|
|
| |
33:Adhesion Alm | Clean electrodes | Electrode adhesion alarm | Clean the electrodes. | |
|
|
|
|
T0652.EPS
IM |