7.2AXFA11 Operation Using a BT200

This section describes procedures for setting parameters using a BRAIN Terminal (BT200). For more details regarding AXFA11 functions, refer to Chapter 6: Parameter Description; and for more details regarding BT200 operation methods, refer to the BT200 User’s Manual (IM IC0A11-01E).

7.2.1BT200 Connection

Connection to a 4 to 20 mA DC signal line

The communication signal of the AXFA11 is superim- posed onto the 4 to 20mA DC analog signals to be transmitted.

AXFA11 terminals

 

 

 

4 to 20mA

Control room

 

 

 

DC signal

 

Receiving

I

Relaying

transmission

Terminal instrument

terminals

line

board

 

I

 

 

 

 

 

 

 

 

 

Receiving

 

 

 

 

resistance:

 

 

 

 

250 to 600 

 

BT200

BT200 BT200

BT200

F0705.EPS

Figure 7.2 Connecting the BT200

IMPORTANT

Restrictions exist with regard to the distance over which communication is possible. (See Chapter 11: Outline.)

IMPORTANT

If the power of flowmeter is turned off within 30 seconds after parameters have been set, these settings will be canceled. Accordingly, please keep the power on for at least 30 seconds after setting parameters.

IMPORTANT

After approximately 5 minutes of inactivity, the Auto Power-Off function will operate to turn your BT200 off.

7. OPERATION VIA BRAIN TERMINAL (BT200)

IMPORTANT

Be sure to set parameters as “Protect” on the write protect function after the finish of param- eter setting. Refer to the “Menu P: Parameter Protection Items” and section “10.2.2” in detail.

NOTE

In case of BT200, the parameters are displayed in English only.

Even if the language with the exception of English is selected at B10/H30: Language, the param- eters are displayed in English upon BT200.

7.2.2The data update and upload/ download function of BT200

(1) The data update of BT200

When the following parameters are displayed, the mea- sured data is updated automatically every seven seconds.

Item

Name (BRAIN)

Item

Name (BRAIN)

 

 

 

 

A10

FLOW RATE (%)

G43

ALM RECORD2

 

 

 

 

A20

FLOW RATE

G44

ALM TIME 2

 

 

 

 

A21

FLW RATE (mA)

G45

ALM RECORD3

 

 

 

 

A30

TOTAL

G46

ALM TIME 3

 

 

 

 

A31

REV TOTAL

G47

ALM RECORD4

 

 

 

 

A32

DIF TOTAL

G48

ALM TIME 4

 

 

 

 

C44

VELOCITY CHK

N20

TEST SI1

 

 

 

 

G22

4-20 BURNOUT

N21

TEST SI2

 

 

 

 

G40

OPERATE TIME

P20

W PROTECT

 

 

 

 

G41

ALM RECORD1

P23

SOFT SEAL

 

 

 

 

G42

ALM TIME 1

 

 

T0703.EPS

(2) Upload/download function of BT200

Upload function is used when the parameters of one AXFA11 are copied to the BT200. And download function is used when the parameters copied to the BT200 are set to another AXFA11.

For details, refer to BT200 User’s Manual (IM 1C0A11-01E). The targeted parameters for upload and download are following.

Item

Name (BRAIN)

Item

Name (BRAIN)

 

 

 

 

C11/B20

FLOW DAMPING

D10/B30

TOTAL UNIT

 

 

 

 

C31

SIZE UNIT

D11/B31

TOTAL SCALE

 

 

 

 

C32

NOMINAL SIZE

E10/B32

PULSE UNIT

 

 

 

 

C40/B21

FLOW UNIT

E11/B33

PULSE SCALE

 

 

 

 

C41/B22

TIME UNIT

H10/B40

DISP SELECT1

 

 

 

 

C42/B23

FLOW SPAN

H11/B41

DISP SELECT2

 

 

 

 

C43/B24

FLOW DECIMAL

H12/B42

DISP SELECT3

T0704.EPS

7-3

IM 01E20C01-01E

Page 70
Image 70
APC AXFA11G AXFA11 Operation Using a BT200, 1 BT200 Connection, Data update and upload/ download function of BT200