[F15: SI1/2 Active Mode] Setting of the active mode for status input

This parameter sets the active mode for the terminals SI1 and SI2. Active modes cannot be set individually for these two terminals.

Setting

Function

 

 

Short Active

When the status input is set to "Short", occurrence

of the selected event will be recognized.

 

Open Active

When the status input is set to "Open", occurrence

of the selected event will be recognized.

 

 

T0628.EPS

Example: When the “0% signal lock” function is selected for SI1 or SI2

Short

Open

Signal lock status

Signal lock status

Short Active

Open Active

 

F0606.EPS

NOTE

Multiple ranges setting

Parameters from F30 to F41 are used with the automatic multiple ranges and the multiple ranges switching via external status input. The followings will describe the setting method for each range.

The multiple ranges use the following parameters:

[B23: Flow Span ] Setting of the flow rate span (Setting of the forward No.1 range)

[F30: Forward Span 2] Setting of the forward No. 2 range

[F31: Forward Span 3] Setting of the forward No. 3 range

[F32: Forward Span 4] Setting of the forward No. 4 range

[F33: Reverse Span 1] Setting of the reverse No. 1 range

[F34: Reverse Span 2] Setting of the reverse No. 2 range

[F35: Reverse Span 3] Setting of the reverse No. 3 range

[F36: Reverse Span 4] Setting of the reverse No. 4 range

6. PARAMETER DESCRIPTION

[F40: Auto Range Hys] Setting of the automatic multiple range hysteresis width

[F41: Bi Direction Hys] Setting of the forward/ reverse flow measurement hysteresis width

Multiple Ranges Setting 1:

Automatic multiple ranges switching

￿When the instantaneous flow rate exceeds 100% of the range, transition to the next range (up to four ranges) is carried out automatically. Furthermore, when the flow is in reverse, the reverse range can also be automatically selected.

￿Range switching can be confirmed according to the status of the SO1 and SO2 status output terminals. Refer to Table 6.4.1: Status Output for Automatic Multiple Ranges Switching for details of status output conditions for each range.

Status Output for Automatic Multiple Ranges Switching

Operations are performed in accordance with the following table when the active mode has been set to “Closed (On) Act” using F14: SO1/2 Active Mode. Operating patterns are reversed when the active mode has been set to “Open (Off) Act”.

Table 6.4.1 Status Output for Automatic Multiple Ranges

Switching

Parameter setting

Function

Status output

F10

F11

SO1

SO2

 

 

No Function

No Function

Forward single range

 

Fwd/Rev

No Function

Auto forward/reverse

Forward

Open

Ranges

1 range (SO1)

Reverse

Closed

No Function

Fwd/Rev

Auto forward/reverse

Forward

Open

Ranges

1 range (SO2)

Reverse

Closed

Auto

No Function

Auto forward

Forward 1 range

Open

2 Ranges

2 ranges (SO1)

Forward 2 range

Closed

No Function

Auto

Auto forward

Forward 1 range

Open

2 Ranges

2 ranges (SO2)

Forward 2 range

Closed

 

 

Auto

Auto

Auto forward

Forward 1 range

Open

Open

Forward 2 range

Closed

Open

3 Ranges

3 Ranges

3 ranges

Forward 3 range

Open

Closed

 

 

 

 

 

Auto forward

Forward 1 range

Open

Open

Auto

Auto

Forward 2 range

Closed

Open

4 ranges

4 Ranges

4 Ranges

Forward 3 range

Open

Closed

 

 

 

 

 

 

 

Forward 4 range

Closed

Closed

 

 

 

Forward 1 range

Open

Open

Fwd/Rev

Auto

Auto forward/reverse Forward 2 range

Open

Closed

Ranges

2 Ranges

2 ranges

Reverse 1 range

Closed

Open

 

 

 

 

 

Reverse 2 range

Closed

Closed

 

 

 

Forward 1 range

Open

Open

Auto

Fwd/Rev

Auto forward/reverse

Forward 2 range

Closed

Open

Reverse 1 range

 

 

2 Ranges

Ranges

2 ranges

Open

Closed

 

 

 

Reverse 2 range

Closed

Closed

*: "No Function" is the default value. Only SO1 or SO2 terminals are used for single or dual ranges; accordingly, the unused terminal remains at the default value (i.e., No Function) and can therefore be used for other functions.

T0629.EPS

6-22

IM 01E20C01-01E

Page 50
Image 50