J31: Power Frequency.

6. PARAMETER DESCRIPTION

[J23: Pulsing Flow] Selection of pulsing flow support In a situation where pulsating flow causes error in the average flow value, due to the application of a plunger pump, this parameter provides functionality whereby calculation is controlled and variations in flow rate are followed.

2600 mm (104 in.), set the fixed frequency (49.00 Hz) in case that either AC or DC power supplies is used for converters.

Set “No” forJ30: Power Synch and “49.00” for

J31: Power Frequency.

Setting

Function

No

Normal

 

 

Yes

Support for pulsing flow

 

 

T0639.EPS

[J24: T/P Damp Select] Setting of damping opera- tion

This parameter is used to select that the flow rate value obtained through damping calculation for totalization and pulse output or the instantaneous flow rate value (no damping) for totalization and pulse output.

Setting

Function

 

 

Damp

Damping

 

 

No Damp

No damping

 

 

T0640.EPS

[J30: Power Synch] Setting of power synchroniza- tion

This parameter sellects whether or not the internal frequency is to be synchronized with that of the power supply.

Setting

Function

 

 

No

Not synchronized

 

 

Yes

Synchronized

 

 

T0641.EPS

Following settings are necessary by power supply and by flow tube size.

Power Supply Code 1

(100 to 240 V AC or 100 to 120 V DC)

 

 

Size 2.5 mm (0.1 in.)

Size 1100 mm (44 in.)

 

 

to 1000 mm (40 in.)

to 2600 mm (104 in.)

 

 

 

 

AC power

J30

Power synchronous (Yes)

Power asynchronous (No)

supply

J31

No setting

49.00 Hz

DC power

J30

Power asynchronous (No)

Power asynchronous (No)

supply

J31

Local commercial power

49.00 Hz

 

 

frequency

 

 

 

 

 

 

 

 

T0642-1.EPS

Power Supply Code 2 (24 V AC/DC)

 

 

 

 

 

 

 

Size 2.5 mm (0.1 in.)

Size 1100 mm (44 in.)

 

 

to 1000 mm (40 in.)

to 2600 mm (104 in.)

 

 

 

 

AC power

J30

Power asynchronous (No)

Power asynchronous (No)

supply

J31

Local commercial power

49.00 Hz

 

 

frequency

 

 

 

 

 

DC power

J30

Power asynchronous (No)

Power asynchronous (No)

supply

J31

Local commercial power

49.00 Hz

 

 

frequency

 

 

 

 

 

 

 

 

T0642-2.EPS

[J40: Memo 1] Setting of memo 1

[J41: Memo 2] Setting of memo 2

[J42: Memo 3] Setting of memo 3

These parameters are used with the memo function, and up to 16 characters can be set for each.

[J31: Power Frequency] Setting of power fre- quency

When “Yes” (i.e., in synchrony) has been selected for J30: Power Synch, this parameter is used to display the power supply frequency. If “No” (i.e., not synchro- nized) has been selected, the power supply frequency is to be specified.

IMPORTANT

In situations where a DC power supply is used for converters, set the local commercial power frequency in area where the converter is in- stalled (size 2.5mm to 1000mm). Set “No” for J30: Power Synch and the local commercial power frequency for J31: Power Frequency.

[J50: Software Rev No] Display of software

revision

This parameter is used to display the software’s revision number.

(9) Menu K: Diagnostic Function Setting items

Menu K contains items that are relevant to the diagno- sis of insulation adhesion to the electrode.

[K10: Adhesion Check] Setting of adhesion

diagnostic function

This parameter selects whether or not the adhesion diagnostic function will be carried out.

IMPORTANT

In situations where AXFA11 is combined with a remote flowtube the size of 1100 mm (44 in.) to

6-32

IM 01E20C01-01E

Page 60
Image 60