6. PARAMETER DESCRIPTION
[J23: Pulsing Flow] Selection of pulsing flow support In a situation where pulsating flow causes error in the average flow value, due to the application of a plunger pump, this parameter provides functionality whereby calculation is controlled and variations in flow rate are followed.
2600 mm (104 in.), set the fixed frequency (49.00 Hz) in case that either AC or DC power supplies is used for converters.
Set “No” forJ30: Power Synch and “49.00” for
J31: Power Frequency.
Setting | Function |
No | Normal |
|
|
Yes | Support for pulsing flow |
|
|
T0639.EPS
[J24: T/P Damp Select] Setting of damping opera- tion
This parameter is used to select that the flow rate value obtained through damping calculation for totalization and pulse output or the instantaneous flow rate value (no damping) for totalization and pulse output.
Setting | Function |
|
|
Damp | Damping |
|
|
No Damp | No damping |
|
|
T0640.EPS
[J30: Power Synch] Setting of power synchroniza- tion
This parameter sellects whether or not the internal frequency is to be synchronized with that of the power supply.
Setting | Function |
|
|
No | Not synchronized |
|
|
Yes | Synchronized |
|
|
T0641.EPS
Following settings are necessary by power supply and by flow tube size.
Power Supply Code 1
(100 to 240 V AC or 100 to 120 V DC)
|
| Size 2.5 mm (0.1 in.) | Size 1100 mm (44 in.) |
|
| to 1000 mm (40 in.) | to 2600 mm (104 in.) |
|
|
|
|
AC power | J30 | Power synchronous (Yes) | Power asynchronous (No) |
supply | J31 | No setting | 49.00 Hz |
DC power | J30 | Power asynchronous (No) | Power asynchronous (No) |
supply | J31 | Local commercial power | 49.00 Hz |
|
| frequency |
|
|
|
|
|
|
|
| |
Power Supply Code 2 (24 V AC/DC) |
| ||
|
|
|
|
|
| Size 2.5 mm (0.1 in.) | Size 1100 mm (44 in.) |
|
| to 1000 mm (40 in.) | to 2600 mm (104 in.) |
|
|
|
|
AC power | J30 | Power asynchronous (No) | Power asynchronous (No) |
supply | J31 | Local commercial power | 49.00 Hz |
|
| frequency |
|
|
|
|
|
DC power | J30 | Power asynchronous (No) | Power asynchronous (No) |
supply | J31 | Local commercial power | 49.00 Hz |
|
| frequency |
|
|
|
|
|
|
|
|
[J40: Memo 1] Setting of memo 1
[J41: Memo 2] Setting of memo 2
[J42: Memo 3] Setting of memo 3
These parameters are used with the memo function, and up to 16 characters can be set for each.
[J31: Power Frequency] Setting of power fre- quency
When “Yes” (i.e., in synchrony) has been selected for J30: Power Synch, this parameter is used to display the power supply frequency. If “No” (i.e., not synchro- nized) has been selected, the power supply frequency is to be specified.
IMPORTANT
In situations where a DC power supply is used for converters, set the local commercial power frequency in area where the converter is in- stalled (size 2.5mm to 1000mm). Set “No” for J30: Power Synch and the local commercial power frequency for J31: Power Frequency.
[J50: Software Rev No] Display of software
revision
This parameter is used to display the software’s revision number.
(9) Menu K: Diagnostic Function Setting items
Menu K contains items that are relevant to the diagno- sis of insulation adhesion to the electrode.
[K10: Adhesion Check] Setting of adhesion
diagnostic function
This parameter selects whether or not the adhesion diagnostic function will be carried out.
IMPORTANT
In situations where AXFA11 is combined with a remote flowtube the size of 1100 mm (44 in.) to
IM