Status Output / Alarm Output

IMPORTANT

Since this is an isolated transistor output, be careful of voltage and polarity when wiring.

Do not apply a voltage larger than 30V DC or a current larger than 0.2A in order to prevent damage to the instrument.

This output cannot switch an AC load. To switch an AC load, an intermediate relay must be inserted as shown in Figure 4.4.13 or Figure 4.4.14.

*The alarm output operates from closed (normal) to open (alarm occurrence) in the default value (as setup upon plant shipment). Changes can be made via the parameter settings.

AXFA11

 

SO1+

Protective diode

 

Load

 

This connection is not possible.

SO2+

 

Load

 

COM

 

External power supply

 

30V DC, 0.2A. max

 

AXFA11

 

SO1+

Relay

Protective

Electromagnetic

diode

valve

 

This connection is not possible.

SO2+

 

 

Electromagnetic

COM

valve

 

External power supply

AC power supply

30V DC, 0.2A. max

 

 

F0421.EPS

Figure 4.4.13 Status Output Connection

4. WIRING

AXFA11

AL+

Protective diode

 

 

 

Load

 

 

AL-

 

 

This connection is not possible.

AXFA11

AL+

Relay

 

 

 

 

 

Electromagnetic

 

AL-

valve

 

 

 

External power supply

AC power supply

 

30V DC, 0.2A. max

 

 

 

F0422.EPS

Figure 4.4.14 Alarm Output Connection

4-9

IM 01E20C01-01E

Page 19
Image 19