10.3.3 Disagreement Between Indication and Actual Flow

10. MAINTENANCE

10.3.3 Disagreement Between Indication and Actual Flow

START

Are parameters

NO

 

 

 

set correctly?

 

Set the parameters correctly.

YES

Examine the condition of the fluid in the flowtube, of bubbles, and of grounding.

Was zero

NO

adjustment carried out

 

correctly?

 

YES

 

Is the tube

NO

filled completely with

 

fluid?

 

YES

 

Does the fluid

YES

contain bubbles?

 

NO

 

Is the

NO

flowmeter completely

 

grounded?

 

YES

 

Is the

 

signal cable

NO

terminal treatment

 

complete?

 

YES

 

Does the

NO

conductivity stay within the

 

specified range?

 

YES

 

Is the

YES

fluid likely to contain

 

deposits?

 

NO

 

Confirm that a valve is located immediately upstream of the flowtube (i.e., on the upstream side).

Is a valve

YES

located immediately upstream

 

of the flowtube?

 

NO

Remove the flowtube and examine the insulation resistance.

Signal terminals A-C and B-C:

NO

100 M or more?

 

Excitation terminals EX1-C:

 

1 M or more?

 

YES

 

Examine the operation of the converter using an AM012 calibrator in order to determine whether the fault lies in the converter or in the flowtube.

Execute zero adjustment when the flowtube is filled completely with fluid and when the fluid is not moving.

Change the mounting position or piping so that the flowtube is

completely filled with fluid.

Switch to vertical mounting.

Install a defoaming device on the flowtube’s upstream side.

Completely ground the converter and flowtube (Refer to the section on grounding in the relevant user’s manual).

Fully terminate the signal cable using PVC tubing or insulation tape, thus insuring that conductors, inner shields, and outer shields do not make contact with each other or with the case.

F1007.EPS Use within the specified conductivity range.

Remove scales. Especially clean the electrodes and their vincinity.

Use a soft brush or cloth to perform scale removal, and make sure that the lining is not damaged.

If any valve located immediately upstream of the flowtube is used in a

half-open condition, the indication may be affected by turbulence in the flow.

Relocate the valve to the downstream side.

If the valve must be located on the upstream side, ensure that the minimum distance to the flowmeter is 10 times the flow-tube diameter; alternatively, use the valve fully open.

Dry the tube fully to restore insulation resistance.

Contact a service center.

F1007.EPS

10-6

IM 01E20C01-01E

Page 108
Image 108