5. BASIC OPERATING PROCEDURES

5.3.2Setting Example for Numeric-Type Data: Flow rate span

This example describes the setting of the flow rate span for the numeric-type parameter

B23: Flow Span from 100 l/min to 120 l/min.

Start:

Major Item

Parameter

Search Mode

Setting mode

 

 

 

 

 

Setting Mode Condition

 

 

 

Setting Mode

 

 

 

 

 

Major item

 

 

P:Protect

Touch the

switch to access B: Easy Setup.

 

 

B:Easy Setup

 

parameter

 

 

C:Basic Setup

 

 

 

NOTE

Sub-item

Parameter

Search Mode

Major item

Sub-item parameter

Major item

Sub-item selection (B)

B:Easy Setup 50:Auto Zero Exe 10:Language

20:Flow Damping

X4

B:Easy Setup 22:Base Time Unit 23:Flow Span

24:Flow Decimal Pnt

The and symbols to the left of the parameters indicate that additional setting items to those being currently displayed may

also be selected. Use the

switch to

cycle through these items.

 

 

 

Sub-item Parameter Search Mode has been accessed in this screen.

Touch the switch to move the cursor to

B23: Flow Span.

The cursor has been moved to B23: Flow Span in this screen. (Sub-item selection screen (B))

Parameter Replacement Mode

Selection of the appropriate parameter

B23:Flow Span

100 l/min

000100. l/min

Rng:0.00001 32000

Touch the switch to access Parameter Replacement Mode.

Once Parameter Replacement Mode has been selected, the digit that can be replaced will be flashed on and off. When in this condition, confirm the relevant setting range as displayed at the bottom of the screen and then set the parameter as required.

In this example, the parameter will be set to “120 l/ min”.

B23:Flow Span

100l/min

0000100. l/min

Rng:0.00001 32000

X4

X2

F0505-1.EPS

NOTE

When setting a new value, use the switch to move from digit to digit, and use the switch to cycle through values for each indi- vidual digit. In addition to digit, it is also possible to select a decimal point, and this allows the position of the decimal point to be changed.

Modify the value to “120 l/min” as follows:

Touch the switch to move the cursor to the position for multiples of 10. Then, touch the switch to change the value at this position from “0” to “2”. When the value of “120” has been setup, touch

the switch.

5-7

IM 01E20C01-01E