(8) Menu J: Auxiliary Function Setting items

This parameter selects the display content of the second line for display unit.

[H12: Display Select3] Setting of the third line for display unit

Refer to the description for parameter B42

This parameter selects the display content of the third line for display unit.

[H20: Display Cycle] Setting of the display cycle This parameter sets the cycle for the display-response

6. PARAMETER DESCRIPTION

value is required for the lower limit.

The indications of the instantaneous flow rates (%, Actual instantaneous flow rate, mA, Bar graph) on the display unit are the same action.

Example: Situation where low limit is set to 10%

Output

5.6mA(10%)

speed of display unit. Settings should be made in accordance with the measurement environment by, for example, setting a longer display cycle when using the equipment in low temperatures.

4mA

NOTE

Manual backgroundInput

F0655.EPS

[H30: Language] Selection of language used for the display unit.

Refer to the description for parameter B10

This parameter can be used to select the language for display unit.

If “2.4mA or less” has been set forG21:4-20mA Alarm Out, 2.4mA or less will be output upon an alarm occurrence, regardless of the low limit setting.

NOTE

(8) Menu J: Auxiliary Function Setting items

Menu J contains setting items such as the flow direc- tion, rate limits, and current output limits.

[J10: 4-20mA Low Cut] Setting of the low-cut range for current output.

This parameter is used to force current output to 0%(i.e., 4mA) in the vicinity of 0% and setting for the current (4 to 20mA) output low cut is made using a percentage of the smallest flow rate span. However, the low cut function will be terminated if this parameter is set to 0%.

The indications of the instantaneous flow rates (%, Actual instantaneous flow rate, mA, Bar graph) on the display unit are the same action.

Example: Situation where low cut is set to 10%

If the setting value for the low limit is not less than the high limit value (as set using J12: 4- 20mA High Lmt), the setting alarm “4-20 Lmt Err” will be displayed.

This parameter has no effect on pulse output or the totalization function.

[J12: 4-20mA High Lmt] Setting of the high limit for current output

This parameter is used to restrict high current portions of current (4 to 20mA) output, and it is initially set to 120%. Setting should be performed when a lower value is required for the higher limit.

The indications of the instantaneous flow rates (%, Actual instantaneous flow rate, mA, Bar graph) on the display unit are the same action.

Output

Setting value

5.6mA(10%)

 

 

4mA

0.5%

0.5%

Input

Setting range: 0 to 10%

 

Hysteresis

 

 

fixed at 1%

 

 

F0615.EPS

[J11: 4-20mA Low Lmt] Setting of the low limit for current output

This parameter is used to restrict low current portions of current (4 to 20mA) output, and it is initially set to -20%. Setting should be performed when a higher

Example: Situation where high limit is set to 90%

Output

18.4mA(90%)

4mA

Input

F0656.EPS

NOTE

If “21.6mA or more” has been set forG21:4-20 mA Alarm Out, 21.6mA or more will be output upon an alarm occurrence, regardless of the high limit setting.

6-30

IM 01E20C01-01E