and “dl” is set forD31: Ttl User Unit.

“dl” is indicated for the totalized units in the Display Mode and is counted in 1 dl steps.

(4) Menu E: Pulse Setting items

Menu E contains items relevant to pulse output.

[E10: Pulse Unit] Setting of the pulse units

Refer to the description of parameter B32: Pulse

Unit

[E11: Pulse Scale] Setting of the pulse scale

Refer to the description of parameter B33: Pulse

Scale

[E12: Pulse Width] Setting of the pulse width This parameter selects the pulse width (i.e., ms : millisecond) that is output.

Data Range

 

Setting

Pulse Rate (pps)

 

 

 

 

Maximum Value

Minimum Value

 

 

 

 

 

 

(0)

50% Duty

11000

0.0001

 

 

 

 

(1)

0.05ms

10000

(pps: pulses per second)

 

 

 

 

(2)

0.1ms

5000

 

 

 

 

 

(3)

0.5ms

1000

 

 

 

 

 

(4)

1ms

500

 

 

 

 

 

(5)

20ms

25

 

 

 

 

 

(6)

33ms

15

 

 

 

 

 

(7)

50ms

10

 

 

 

 

 

(8)

100ms

5

 

 

 

 

 

T0621.EPS

6. PARAMETER DESCRIPTION

[E13: Pulse Low Cut] Setting of the pulse output stop range

This parameter allows the settings to be made which prevent pulse output when the flow rate is at or below the low-cut setting value. In cases where there are multiple ranges or forward/reverse ranges, low cut is carried out at the setting value for the smallest span (i.e., an integer between 0 and 10%).

Example: When the first range = 10 m3/h, the second range = 100 m3/h, and the Pulse Low Cut = 3%, no pulse output is carried out at flow rates of 0.3 m3/h or lower.

[E20: Pulse Active Mode] Setting of active mode for pulse output

This parameter sets whether pulse-output active mode is to be “On Active” or “Off Active”.

Item

Description

 

 

Closed (On) Act

When pulse output is carried out, contacts

 

are Closed (On).

 

 

Open (Off) Act

When pulse output is carried out, contacts

 

are Open (Off).

 

 

 

T0622.EPS

Example: When E12: Pulse Width is 1 ms, output takes place as follows in accordance with the

setting for E20: Pulse Active Mode.

Closed(On)

Open(Off)

1 ms

1 ms

Closed (On) Active

Open (Off) Active

 

F0604.EPS

NOTE

*:The pulse width with the exception of “50% Duty” is the “Closed (On)” time for each pulse in case that “Closed (On) Act” is sellected for

E20: Pulse Active Mode.

Closed(On)

Open(Off)

Pulse width

F0603.EPS

A limit applies to the maximum pulse scale that can be set with respect to the pulse width. If a value in excess of this limit is set, a setting alarm will be displayed.

6-19

IM 01E20C01-01E