and “dl” is set forD31: Ttl User Unit.
“dl” is indicated for the totalized units in the Display Mode and is counted in 1 dl steps.
(4) Menu E: Pulse Setting items
Menu E contains items relevant to pulse output.
[E10: Pulse Unit] Setting of the pulse units
→Refer to the description of parameter B32: Pulse
Unit
[E11: Pulse Scale] Setting of the pulse scale
→Refer to the description of parameter B33: Pulse
Scale
[E12: Pulse Width] Setting of the pulse width This parameter selects the pulse width (i.e., ms : millisecond) that is output.
Data Range
| Setting | Pulse Rate (pps) | |
|
|
| |
| Maximum Value | Minimum Value | |
|
| ||
|
|
|
|
(0) | 50% Duty | 11000 | 0.0001 |
|
|
|
|
(1) | 0.05ms | 10000 | (pps: pulses per second) |
|
|
|
|
(2) | 0.1ms | 5000 |
|
|
|
|
|
(3) | 0.5ms | 1000 |
|
|
|
|
|
(4) | 1ms | 500 |
|
|
|
|
|
(5) | 20ms | 25 |
|
|
|
|
|
(6) | 33ms | 15 |
|
|
|
|
|
(7) | 50ms | 10 |
|
|
|
|
|
(8) | 100ms | 5 |
|
|
|
|
|
T0621.EPS
6. PARAMETER DESCRIPTION
[E13: Pulse Low Cut] Setting of the pulse output stop range
This parameter allows the settings to be made which prevent pulse output when the flow rate is at or below the
Example: When the first range = 10 m3/h, the second range = 100 m3/h, and the Pulse Low Cut = 3%, no pulse output is carried out at flow rates of 0.3 m3/h or lower.
[E20: Pulse Active Mode] Setting of active mode for pulse output
This parameter sets whether
Item | Description |
|
|
Closed (On) Act | When pulse output is carried out, contacts |
| are Closed (On). |
|
|
Open (Off) Act | When pulse output is carried out, contacts |
| are Open (Off). |
|
|
| T0622.EPS |
Example: When E12: Pulse Width is 1 ms, output takes place as follows in accordance with the
setting for E20: Pulse Active Mode.
Closed(On)
Open(Off)
1 ms | 1 ms |
Closed (On) Active | Open (Off) Active |
| F0604.EPS |
NOTE
*:The pulse width with the exception of “50% Duty” is the “Closed (On)” time for each pulse in case that “Closed (On) Act” is sellected for
E20: Pulse Active Mode.
Closed(On)
Open(Off)
Pulse width | F0603.EPS |
A limit applies to the maximum pulse scale that can be set with respect to the pulse width. If a value in excess of this limit is set, a setting alarm will be displayed.
IM |