| 6. PARAMETER DESCRIPTION | |
|
|
|
|
|
|
|
|
|
| Name |
| Data range |
| Position | Default value |
|
|
|
|
|
| Units |
|
| ||
Item | Display unit | R/W | Display unit | of decimal | Description |
| ||
(*): Indicated item |
| |||||||
| (BRAIN) |
| /BRAIN |
| point |
|
|
|
F12 | SI1 Function | W | No Function |
|
| No Function | Selects function for the SI1 |
|
| (SI1 FUNCTION) |
| 0% Singal Lock |
|
|
| terminal |
|
|
|
| Ext Auto Zero |
|
|
|
|
|
|
|
| Ext Ttl Preset |
|
|
|
|
|
|
|
| Ext Rev Ttl Set |
|
|
|
|
|
|
|
| Ext 2 Ranges |
|
|
|
|
|
|
|
| Ext 3 Ranges |
|
|
|
|
|
|
|
| Ext 4 Ranges |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
F13 | SI2 Function | W | No Function |
|
| No Function | Selects function for the SI2 |
|
| (SI2 FUNCTION) |
| 0% Singal Lock |
|
|
| terminal |
|
|
|
| Ext Auto Zero |
|
|
|
|
|
|
|
| Ext Ttl Preset |
|
|
|
|
|
|
|
| Ext Rev Ttl Set |
|
|
|
|
|
|
|
| Ext 2 Ranges |
|
|
|
|
|
|
|
| Ext 3 Ranges |
|
|
|
|
|
|
|
| Ext 4 Ranges |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
F14 | SO1/2 Active Mode | W | Closed(On) Act |
|
| Closed(On) Act | Selects whether SO1/SO2 output |
|
| (SO ACT MODE) |
| Open(Off) Act |
|
|
| will be set to “On Active” or |
|
|
|
|
|
|
|
| “Off Active.” |
|
|
|
|
|
|
|
|
|
|
F15 | SI1/2 Active Mode | W | Short Active |
|
| Short Active | Selects whether SI1/SI2 input |
|
| (SI ACT MODE) |
| Open Active |
|
|
| will be set to “Short Active” or |
|
|
|
|
|
|
|
| “Open Active.” |
|
|
|
|
|
|
|
|
|
|
F30 | Forward Span2 | W | 0.0001 to 32000 | C40 | 0 to 4 | 1 | Sets flow rate span for forward |
|
| (FWD SPAN2) |
|
| /C41 |
|
| No. 2 range |
|
F31 | Forward Span3 | W | 0.0001 to 32000 | C40 | 0 to 4 | 1 | Sets flow rate span for forward |
|
| (FWD SPAN3) |
|
| /C41 |
|
| No. 3 range |
|
F32 | Forward Span4 | W | 0.0001 to 32000 | C40 | 0 to 4 | 1 | Sets flow rate span for forward |
|
| (FWD SPAN4) |
|
| /C41 |
|
| No. 4 range |
|
|
|
|
|
|
|
|
|
|
F33 | Reverse Span1 | W | 0.0001 to 32000 | C40 | 0 to 4 | 1 | Sets flow rate span for reverse |
|
| (REV SPAN1) |
|
| /C41 |
|
| No. 1 range |
|
F34 | Reverse Span2 | W | 0.0001 to 32000 | C40 | 0 to 4 | 1 | Sets flow rate span for reverse |
|
| (REV SPAN2) |
|
| /C41 |
|
| No. 2 range |
|
F35 | Reverse Span3 | W | 0.0001 to 32000 | C40 | 0 to 4 | 1 | Sets flow rate span for reverse |
|
| (REV SPAN3) |
|
| /C41 |
|
| No. 3 range |
|
F36 | Reverse Span4 | W | 0.0001 to 32000 | C40 | 0 to 4 | 1 | Sets flow rate span for reverse |
|
| (REV SPAN4) |
|
| /C41 |
|
| No. 4 range |
|
F40 | Auto Range Hys | W | 0 to 15 | % | 0 | 10 % | Sets hysteresis width for |
|
| (AUTO RNG HYS) |
|
|
|
|
| automatic range switching |
|
F41 | Bi Direction Hys | W | 0 to 8 | % | 0 | 2 % | Sets hysteresis width for |
|
| (BI DIREC HYS) |
|
|
|
|
| forward/reverse switching |
|
F60 | — | R | Good |
|
|
| See “6.5 Alarm Function”. |
|
| (SELF CHECK) |
| Error |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
(7) Item G (Menu G): Alarm Setting items
Menu G contains setting items relevant to alarm output, burnout, alarm record, etc.
| Name |
| Data range |
| Position | Default value |
| |
Item |
|
|
| Units | Description | |||
Display unit | R/W | Display unit | of decimal | |||||
(*): Indicated item | ||||||||
| (BRAIN) |
| /BRAIN |
| point |
|
| |
G00 | Alarm |
|
|
|
|
|
| |
| (ALARM) |
|
|
|
|
|
| |
|
|
|
|
|
|
|
| |
G10 | Low Alarm | W | % | 0 | Sets level setting value for low | |||
| (LOW ALARM) |
|
|
|
|
| flow rate limit (L) | |
|
|
|
|
|
|
|
| |
G11 | High Alarm | W | % | 0 | 110 | Sets level setting value for high | ||
| (HIGH ALARM) |
|
|
|
|
| flow rate limit (H) | |
G12 | Low Low Alarm | W | % | 0 | Sets level setting value for low- | |||
| (LO LO ALARM) |
|
|
|
|
| low flow rate limit (LL) | |
G13 | High High Alarm | W | % | 0 | 110 | Sets level setting value for high- | ||
| (HI HI ALARM) |
|
|
|
|
| high flow rate limit (HH) |
IM |